ABB AP300 User Manual page 11

Combination ph/redox (orp) sensors
Table of Contents

Advertisement

3. Measure and record the following distances –
see Fig. 2.5:
— outside of pipe/vessel to the reducer at the back end
A
of the valve
.
— length of compression fitting after insertion into the
B
?
reducer
— pipe/vessel wall thickness
— sensor insertion depth
4. Add together measurements from step 3 and, measuring
from the tip of the sensor, mark off total onto the shaft of
the sensor.
Fig. 2.5 Installing the AP304/5 – Installation Measurements
C
.
D
.
B
5. Insert the sensor into the ball valve. Tighten onto the ball
valve on reducer.
6. Open the ball valve fully and gently push the sensor into
the valve until the mark made at 4 reaches the
compression fitting.
7. Tighten the outer compression joint (manually or using a
wrench depending on type of fitting chosen).
8. A mark can also be made distance B from the tip of the
probe. This can then be used to tell when the valve can be
closed when retracting the probe.
D
A
AP300 | Combination pH/Redox (ORP) sensors | IM/AP300 Rev. J
C
C
11

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents