16.2. Basic Functions

Function No.
Setting range
f001
0000 to 1111
f002
0 to 11
f003
0 to 4
f004
0000 to 1111

16.3. Comparator

Function No.
Setting range
f101
-9999 to 9999
f102
-9999 to 9999
f103
0 to 2
f104
-9999 to 9999
f105
1 to 3
f106
0.0 to 5.0
f107
00 to 99
Description
Disable keys:
0: Enabled
1: Disabled
Digital filter:
0: Off
6:0.5 Hz
1:2.5 Hz
7:0.35 Hz
2:2.0 Hz
8:0.25 Hz
3:1.5 Hz
9:0.20 Hz
4:1.0 Hz
10:0.15 Hz
5:0.7 Hz
11:0.10 Hz
Hold mode:
0: Off
1: Sample hold
2: Peak hold
3: Bottom hold
4: Bipolar peak hold
LATCH function:
0: Off
1: On
Description
Upper limit value of comparator.
Lower limit value of comparator.
Comparator mode:
0:Off
1:On when the measured value is not
near zero
2:Always on
Set
the
near
zero
comparator mode.
Hysteresis direction:
1:Upward 2-level judgment
2:Upper / lower limit judgment
3:Downward 2-level judgment
Hysteresis time. (Unit: 0.1 second)
When 0. 0 , the hysteresis is not used.
Hysteresis width. (d)
When 00, the hysteresis is not used.
range
for
the
49
Default value User setting
0000
(Binary)
4
1
0000
(Binary)
Default value User setting
0000
0000
2
0000
2
0. 0
99