12. COMPARATOR FUNCTION

The comparator function compares the measured value against the set values, and the
results are displayed by the LEDs. If the relay option is installed, it outputs the
comparison results (HI, OK or LO) from the rear panel relay output terminals.

12.1. Detailed description of the comparator mode

The relation between the comparison outputs and the upper and lower limit setting
values is as shown below:
Comparison result
HI
OK
LO
* HI is output with a positive over limit, LO is output with a negative over limit.
Upper limit and lower limit values can be negative.
For example, if the upper limit value is -1000 and the lower limit value is -2000, HI is
output for the measured value of -500 and LO is output for the measured value of -2500.
Make sure that the upper limit value is greater than the lower limit value.

12.2. Setting the upper and lower limit value

Set the upper limit value (f101) and lower limit value (f102) of the function using the
following procedure. In the measurement mode, press the
the setting mode for the upper or lower limit value.
In the setting mode for the upper or lower limit value, the AD-4530 blinks the
LED. (The relay output is not affected.)
The setting value is displayed. Set the value by operating the key.
Shift the blinking digit to the value to be changed.
Change the value of the blinking digit.
Change the polarity.
Store the setting, and return to the measurement mode.
Cancel the setting, and return to the measurement mode.
Comparison condition
Upper limit < Measured value
Lower limit  Measured value  Upper limit
Measured value < Lower limit