Calibration; Calibration Modes - AND AD-4530 Instruction Manual

Digital indicator
Table of Contents

Advertisement

The AD-4530 measures voltage signals from sensors and displays the values.
Calibration is performed so that the AD-4530 performs correctly.
The decimal point (Cf01), minimum scale value (Cf02) and rated capacity (Cf03) are
set using the function mode.
The zero point input voltage (Cf04), the span input voltage (Cf05) and the display value
for the span input voltage (Cf06) are set using the calibration mode.
Calibration setting by the function mode is also available. (Digital calibration)
* During calibration, maintain a stable environment to prevent calibration errors.
* You can confirm stabilization by confirming that the
* The decimal point blinks to indicate that a number is being displayed, not a
measured value.

7.1. Calibration Modes

In the measurement mode, press the
Enter the zero point calibration mode.
Return to the measurement mode.
7.1.1. Zero point calibration mode
With nothing on the load cell, wait for it to stabilize, and then
press the
Perform zero point calibration, and
proceed to the span calibration mode.
Cancel zero point calibration, and
proceed to the span calibration mode.
Hold to display the mV/V of the zero point.

7. CALIBRATION

key.
HOLD
+
keys to enter the calibration mode.
18
LED is on.
Span calibration