8.3.5. Serial Communication
Function No.
Function
Setting range
f301
Baud rate
2.4 to 38.4
f302
Data bit
7 to 8
length
f303
Parity
0 to 2
f304
Stop bit
1 to 2
f305
Terminator
1 to 2
f306
Communication
1 to 2
mode
f307
Model No.
00 to 99
f401
Unit character1
f402
Unit character2
f403
Unit character3
f404
Unit character4
f405
Unit character5
00 to 7F
8.3.6. Error
Display
The data cannot be acquired from
ad?e
A/D converter.
Correct data cannot be read from
eepe
EEPROM.
Cale
Calibration data error.
C e*
Calibration error.
A setting value is out of range.
dt?e
Description
2. 4 : 2400 bps
4. 8 : 4800 bps
9. 6 : 9600 bps
19. 2 :19200 bps
38. 4 :38400 bps
7:7 bits
8:8 bits
0:None
1: Odd
2: Even
1:1 bit
2:2 bits
1:CRLF
2:CR
1: Stream
2: Manual print
ID that is added to the serial output.
When setting is 00, the ID is not added.
Unit character added to serial output.
Set with the hexadecimal ASCII code.
All characters after 00 are ignored.
Cause
26
Remedy
Repair is required.
Perform initialization.
If the initialization does not clear the
error, repair is required.
Perform calibration.
Refer to "7.2.Calibration Errors".
Check setting values and correct if
necessary.
Default value
Setting type
2. 4
P
7
P
2
P
1
P
1
P
2
P
00
D
00
00
00
00
00
(Hexadecimal)