ADInstruments PowerLab Owner's Manual page 28

Galvanic skin response amplifier
Table of Contents

Advertisement

You are using an early version of LabChart.
The BNC or I
The front-end is faulty.
On starting up the soft ware, an alert indicates that there is a problem with the front-end or
driver
The correct BP/GSR drivers are not installed on your computer.
You are using an early version of LabChart.
The BNC or I
The front-end is faulty.
Some soft ware settings don't resemble those in this guide
You are using an early version of the front-end driver, or of LabChart. Some changes
may have been made since then.
28
• Turn everything off . Check to see that all cables are firmly seated and screwed
in. BNC cables from the GSR Amp must be connected to a positive input on the
PowerLab. Make sure the input is the same channel from which you expect to use
the front-end in the soft ware. Aft er checking connections, restart the PowerLab
and application to see if this has fixed the problem.
• Upgrade to the latest version of the soft ware. Contact your ADInstruments
representative for information.
2
C cable is faulty.
• Replace the cable and try again. Immediately label all cables proved faulty so
that you don't use them again by accident.
• This is the least likely event. If the front-end will not work properly aft er the
previous measures, then try using it on another PowerLab. If the same problems
recur with a second PowerLab, the front-end may be faulty. Contact your
ADInstruments representative to arrange for repairs.
• Reinstall the soft ware.
• Upgrade to the latest version of the soft ware. Contact your ADInstruments
representative for information.
2
C cable is faulty.
• Replace the cable and try again. Immediately label all cables proved faulty so
that you don't use them again by accident.
• This is the least likely event. If the front-end will not work properly aft er the
previous measures, then try using it on another PowerLab. If the same problems
recur with a second PowerLab, the front-end may be faulty. Contact your
ADInstruments representative to arrange for repairs.
• Upgrade to the latest version of the soft ware. Contact your ADInstruments
representative for information.
Owner's Guide