Test Reports
• After all required microtiter plate readings have been collected, Data
Manager will display a Test Report (see screen image on page 21). Reports
may also be automatically exported by checking the box in the Options
menu.
• The Test Report can be exported in HTML or XML format and may be
imported to a spreadsheet program for further analysis.
• The yellow or orange colored cells represent dispensed volumes or dilution
ratios that exceed the user-defined limits for Relative Inaccuracy and/or
Coefficient of Variation.
• If the Heat Map was enabled in the Plate Layout, the Test Report will
display colored data points representing the variation in volumes within
that Layout Group.
• If the Security System is active, Data Manager will prompt for a password
once all scans are completed. This process attaches an electronic signature
to the end of the Test Report and logs the action in the Audit Trail.
•
The electronic signature should be considered the equivalent of a dated,
handwritten signature applied to the test result.
• Comments that are added to the report are saved, but any changes to the
way data is viewed, such as Change Limits, will not be saved.