Page 1
To ensure safe and proper use of the balance, please read this manual carefully. • After reading this manual, store it in a safe place near the balance, so you can review it as needed. SHINKO DENSHI CO., LTD. 410026M31...
Page 3
Preface Thank you very much for having purchased our dust- and water-proof intrinsic safety explosion-proof structure electronic scale. This document is the Operation Manual for the following dust- and water-proof intrinsic safety explosion-proof structure electronic scale. In the first place, install this product properly referring to the Installation Manual attached separately, and then read this document.
Page 4
Instructions ● The copyright of this document belongs to SHINKO DENSHI CO., LTD. Reprinting or duplicating of all or part of this document without notice shall not be allowed. ● Please note that product improvement or modification may cause partial discrepancy between the product and the description of this document.
● The copyright of this document is held and reserved by SHINKO DENSHI CO., LTD. Duplicating or disclosing its drawings and engineering materials without prior approval of SHINKO DENSHI CO., LTD. in writing is not permitted.
How to use this document ■ Symbols used in this document Understand the meaning of the following symbols and observe the instructions of this document. Symbols Meaning Used for the situation that invites an imminent risk of death or severe injury unless avoided.
Page 7
■About how to read this document This document consists of the following contents: Describes about operating precautions, names 1 When beginning to use and functions of each section, etc. Please be sure to read this section when using this product for the first time.
Contents Preface ..........................i Important Notice ........................iii How to use this document ....................iv Contents ..........................vi 1 Prior to use ........................1 Precautions ........................1 Names and functions of each section ................3 Performance of operation keys ..................4 How to interpret the display .................... 5 1-4-1 Main LCD ........................
Page 9
Stability discrimination width ..................30 Stability discrimination frequency ................. 31 Response speed ......................31 Weight renewal interval ....................32 5 User information setting ....................33 Hierarchy of user information setting ................33 Measurer’s ID setting ....................33 Product name ID setting ....................34 Lot number setting ......................
Page 10
Span adjustment history ....................59 Scale ID setting ......................60 Maintenance setting ..................... 60 Date and time setting ....................61 Date indication format ....................61 Output character setting ....................61 Password control ......................62 8-10 Password change ......................62 8-11 Password cancellation history ..................
FZ- Ex series operetion manual 1 Prior to use 1 Prior to use Precautions ■ No disassembling or modification. Unless specifically stated in this document, disassembling or modification of this product, mounting or removal of an undesignated component no longer maintains the function of the explosion-proof structure, leading to a serious accident or bodily injury.
Page 12
1 Prior to use FZ-Ex series operetion manual ■ Do not use the product in an abnormal condition. If it should happen that an abnormal event such as smoking or unusual odor occurs, ask the store where you purchased the product for repair. Keeping using the product may result in an electric shock or fire.
1 Prior to use FZ-Ex series operetion manual Performance of operation keys Type / name of a key Performance [On / Off] Turns on and off the power for the scale. [Direction] Used for function setting. [Transfer] Used for outputting. [Function F] Used for function calling.
FZ- Ex series operetion manual 1 Prior to use How to interpret the display 1-4-1 Main LCD Name Description Symbol Gram Represents gram unit. Kilogram Represents kilogram unit. Percent Lights when in the percent scale mode. Zero point Indicates the zero point. Plus Plus Minus...
1 Prior to use FZ-Ex series operetion manual 1-4-2 Sub LCDs ( i03 only) ■Upper sub LCD ■Lower sub LCD Name Description Symbol Gram Represents gram unit. Kilogram Represents kilogram unit. Percent Lights when in the percent scale mode. 7-segment display Displays numbers and simple letters.
FZ-Ex series operetion manual 2 Basic usage 2 Basic usage Turning on / off the power, and checking for the operation Turn on and off the power for this product. Turn on the power . Turn on the power for the power supply box.
2 Basic usage FZ-Ex series operetion manual Turn off the power for the scale. Push and hold the [On/Off] key. The product becomes standby status and the symbol ( ) lights. Pushing and holding the [On/Off] key obtains the standby status from any operation status.
Zero adjustment range when in use Zero adjustment range when in use is limited in this product. The available zero adjustment range when in use is shown below: Model Lower limit (g) Upper limit (g) -9.3 FZ623Ex -48 FZ3202Ex -93 FZ6202Ex -225 FZ15001Ex -450...
2 Basic usage FZ-Ex series operetion manual Weighing an object placed on a container (tare) When weighing an object to be weighed with the object placed on a container (tare), the weight of the container must be subtracted from the total weight to get the actual weight of the object to be weighed.
FZ-Ex series operetion manual 2 Basic usage Weighing with an object to be weighed added Place an added object to be weighed and weigh the weight of the added object. Performing the tare weight subtraction with the object to be weighed which has been already weighed makes it possible to weigh the mass of a next object to be weighed with the previous object to be weighed remaining placed.
2 Basic usage FZ-Ex series operetion manual Selecting the main LCD indication The main LCD and the sub LCD can be used in combination with each other. The content of main LCD indication changes in the following sequence: In the case the percent scale function and the adding function are set, selection (switching) of the main LCD indication is available.
FZ-Ex series operetion manual 2 Basic usage Basic operation The menu of this product is divided into two as described below: (1)Setting menu The menu to set a variety of functions (2)Execution menu The menu not to set but only to execute the program. 2-7-1 Hierarchy of a setting menu The setting menu of this product is divided into four, from the first layer to the third layer and...
2 Basic usage FZ-Ex series operetion manual 2-7-2 Operation of the setting menu, setting of various functions To perform settings for various functions from the state of weighing, chiefly execute the following procedure. (1)Push [Function F] key to enter respective setting from the state of weighing. (2)Shift to the intended setting item using the [Direction] key.
Page 25
FZ-Ex series operetion manual 2 Basic usage To return to the state of weighing after setting various functions, chiefly execute the following procedure. (1) Push the [Shift] and then [Clear Home] keys at any of the first, second or third layer. -15-...
2 Basic usage FZ-Ex series operetion manual 2-7-3 Operation of the setting menu, inputting of numeric values Numeric value inputting is limited to seven digits at a maximum. ■Example) When inputting 12345 Input “1”. Push the [Numeric keypad] “1.” Number “1” is displayed on the extreme right like 「...
FZ-Ex series operetion manual 2 Basic usage 2-7-4 Operation of the setting menu, inputting of characters ■Operation of character input Push the [Numeric keypad] “ 2 ” Push the [Numeric keypad] “ 2 ”. Number “2”is displayed on the extreme 」.
Page 28
2 Basic usage FZ-Ex series operetion manual Input “ C ” Push “Right” of the [Direction] key. The digit that has been input moves to the right. Keep pushing the [Numeric keypad] “2” till the indication changes to“ 「ABC 」 ”. Discontinue inputting.
3 Functions related to the operation FZ-Ex series operetion manual 3 Functions related to the operation Settings to change the scale operations Hierarchy of functions related to the operation ” “ can not be used. -19-...
3 Functions related to the operation FZ-Ex series operetion manual Unit setting Unit in the weight mode can be set either in 「g」 or 「kg」. Select a setting menu. Select the unit setting. (“2-7-2 Operation of the setting menu, setting of various functions”...
3 Functions related to the operation FZ-Ex series operetion manual Adding function Weighs a plurality of objects to be weighed in sequence and indicates its total value. The adding function includes two ways of calculating method. Method of weighing objects to be weighed while replacing the objects Addition accumulating function Method of weighing objects to be weighed without replacing the Net adding function...
3 Functions related to the operation FZ-Ex series operetion manual 3-4-1 Weighing by means of the plus side addition Place a first object to be weighed. After 「 」 appears, push [Enter] key. The weighed value is stored and 「 」...
3 Functions related to the operation FZ-Ex series operetion manual 3-4-2 Weighing by means of the minus side addition Place an object to be weighed and perform the tare weight subtraction. After 「 Tare 」 appears, push [Enter] key. The weighed value is stored and 「 」...
3 Functions related to the operation FZ-Ex series operetion manual Comparator function It is possible to preset threshold values and determine whether or not a measured value is within the range defined by the preset values. The comparator function can be used in any scale mode, i.e. weight scale mode, percent scale mode, and coefficient scale mode.
3 Functions related to the operation FZ-Ex series operetion manual 3-5-3 Comparator function setting Select a setting menu. Select comparator function. (“2-7-2 Operation of the setting menu, setting of various functions” and “3-1 Hierarchy of functions related to the operation”) Set comparator function.
3 Functions related to the operation FZ-Ex series operetion manual Buzzer setting This is a convenient function for key inputting and use of the comparator function. Select a setting menu. Select the buzzer setting function. (“2-7-2 Operation of the setting menu, setting of various functions”...
3 Functions related to the operation FZ-Ex series operetion manual Conditions for stability waiting Can not be used. Set when to indicate the weighed value after the zero adjustment or tare weight subtraction; either after or before the weighed value stabilizes. Select a setting menu.
3 Functions related to the operation FZ-Ex series operetion manual 3-10 Direct start Setting to the direct start makes it possible to turn on and off the power with the switch on the power supply box without pushing the [On/Off] key. Select a setting menu.
FZ-Ex series operetion manual 4 Function srelated to the performance Function srelated to the performance Set the scale indication stability and response speed. Hierarchy of functions related to the performance ” “ can not be used. -29-...
4 Function srelated to the performance FZ-Ex series operetion manual Zero tracking Setting to the zero tracking function makes it possible to automatically correct the zero point fluctuation caused by the temperature fluctuation, etc. that is likely to occur when “0” is indicated, through which the “0”...
FZ-Ex series operetion manual 4 Function srelated to the performance Stability discrimination frequency Can not be used. The larger numeric value is set, the higher stability is obtained. Select a setting menu. Select the stability discrimination frequency. (“2-7-2 Operation of the setting menu, setting of various functions”...
4 Function srelated to the performance FZ-Ex series operetion manual Weight renewal interval Can not be used. This is a function to output data at regular intervals. Select a setting menu. Select the sampling time. (“2-7-2 Operation of the setting menu, setting of various functions”...
FZ-Ex series operetion manual 5 User information setting 5 User information setting Set various user IDs and upper and lower limit values. Hierarchy of user information setting (1) It is possible to register a number for an ID up to a maximum of ten digits. (2) Up to 100 IDs that have been set can be registered individually from “001”...
5 User information setting FZ-Ex series operetion manual Product name ID setting An ID can be provided for each product name. Select a setting menu. Select the product name ID. (“2-7-2 Operation of the setting menu, setting of various functions” and “5-1 Hierarchy of user information setting”) Set the product name ID.
FZ-Ex series operetion manual 5 User information setting Preset tare weight setting Inputting, registration and calling of a preset tare weight value can be performed. 5-6-1 Inputting of a preset tare weight value There are two ways of inputting a reference value and upper and lower limit values as described below: - Numeric value setting method: Inputting a setting value directly via [Numeric keypad] operation...
Page 46
5 User information setting FZ-Ex series operetion manual The preset tare weight setting can be performed out of the scale mode as a shortcut according to the following procedure: Set a tare weight value. [Numeric value setting method] Push the [Preset tare] key in the scale mode.
FZ-Ex series operetion manual 5 User information setting 5-6-2 Registration of a preset tare weight value Nine preset tare weight values can be registered. Set a preset tare weight value. (“2-7-2 Operation of the setting menu, setting of various functions” and “5-1 Hierarchy of user information setting”) Register the preset tare weight value.
5 User information setting FZ-Ex series operetion manual Setting of a percent scale reference value The weight of an object to be weighed is indicated in percent relative to the reference weight. Set the reference weight by way of either the numeric value setting method, which requires the inputting of a numeric value, or the actual value setting method, which requires the weighing of a sample.
FZ-Ex series operetion manual 5 User information setting An indication of “L-Err” signifies that the reference weight is below the limit weight and that the weight is unmeasurable. Percent scale limit weight FZ623Ex 0.1 g FZ3202Ex、6202Ex FZ15001Ex FZ30K0.1GEx、FZ60K0.1GEx FZ100K1GEx、FZ200K1GEx 100 g FZ150K1GFEx、FZ300K1GFEx...
Page 50
5 User information setting FZ-Ex series operetion manual Select the upper limit value setting. Push the [Direction] key. Select “37.TS.” Push the [Direction] key. Select “372.HI.” Push the [Enter] key. Set an upper limit value. Input an upper limit value with [Numeric keypad].
Page 51
FZ-Ex series operetion manual 5 User information setting The discrimination value setting of the comparator function can be performed out of the scale mode as a shortcut according to the following procedure: Select the reference value setting. (In the case of the relative value discrimination) Push the [Target] key in the scale mode.
5 User information setting FZ-Ex series operetion manual 5-8-2 Actual value setting method Select a setting menu. Select the discrimination value setting of the comparator function. (“2-7-2 Operation of the setting menu, setting of various functions”) Select the reference value setting. (In the case of the relative value discrimination) Push the [Direction] key.
Page 53
FZ-Ex series operetion manual 5 User information setting Set a lower limit value. Push the [Function F] key. Place an object to be weighed that is equivalent to the lower limit value. Push the [Enter] key. The lower limit value is stored. The operation mode returns to the scale mode.
Page 54
5 User information setting FZ-Ex series operetion manual The discrimination value setting of the comparator function can be performed out of the scale mode as a shortcut according to the following procedure: Select the reference value setting. (In the case of the relative value discrimination) Push the [Target] key in the scale mode.
FZ-Ex series operetion manual 5 User information setting Coefficient value setting Can not be used. The value that is obtained by multiplying a measured weight by a predetermined coefficient can be indicated. For example, in the case the coefficient is “2.35” and the measured weight is “2000 g”, a value of “4700 g”...
6 External input / output functions FZ-Ex series operetion manual 6 External input/output functions This function is used for communication through the external peripheral devices. Hierarchy of the external input/output functions Can not be used. -46-...
FZ-Ex series operetion manual 6 External input / output functions Connecter terminal numbers and their functions Input/output to and from an external device such as a personal computer via the RS-232C is available. The RS-232C interface for this product is the D-SUB9P type. The RS-232C connector pin alignment for this product is as shown below: Terminal no Signal name...
FZ-Ex series operetion manual 6 External input / output functions 6-4-4 Input command composition Composed of four characters including a terminator (CR=0DH / LF=0AH). 6-4-5 Transmission procedure Send an input command from an external device to the scale. Since transmission and reception are performed by way of full-duplex communication method, the input command can be transmitted irrespective of the transmission timing from the scale.
6 External input / output functions FZ-Ex series operetion manual 6-4-6 Command format Please take care not to take alphabetical “O” for Arabic number “0.” Response Code Code Description A00, Exx ACK, NAK (C1) (C2) format format 0x20 Tare weight subtraction (SP) 0x54 A00:...
FZ-Ex series operetion manual 6 External input / output functions GZII format This is different from “6-4 GZIII format” only in the operation of the T-command. In the GZII format, the tare weight subtraction / zero adjustment is executed by the T-command. For other specifications, please refer to “6-4 GZIII format”.
6 External input / output functions FZ-Ex series operetion manual Response 6-6-1 Response command format (when set to the A00, Exx format) Composed of five characters including a terminator (CR=0DH / LF=0AH) 6-6-2 Response command Code(A1) Code(A2) Code(A3) Description Normal completion ※Command error (Abnormal command received) 0 - 9...
FZ-Ex series operetion manual 6 External input / output functions Power supply box communication setting Perform the power supply box communication setting in line with the peripheral device to which the output is transmitted. Select a setting menu. Select the power supply box communication setting.
Page 64
6 External input / output functions FZ-Ex series operetion manual Set the output conditions. Push the [Direction] key. Select “413.OC.”. Input a setting value. 「413.OC. 0」 : Output stop 「413.OC. 1」 : Continuous output at all Inputting of the setting value times 「413.OC.
FZ-Ex series operetion manual 6 External input / output functions Set the stop bit. Push the [Direction] key. Select “418.ST.”. Input a setting value. 「418.ST. 1」 :1 bit 「418.ST. 2」 :2 bits Inputting of the setting value Set unused high order digit. Push the [Direction] key.
Functions related to the lock FZ-Ex series operetion manual 7 Functions related to the lock Perform the setting for the prohibition of change of menu items and the disabling of key operation, etc. Hierarchy of functions related to the lock Locking of functions related to the operation Various setting menus can be locked.
FZ-Ex series operetion manual Functions related to the lock Set the user information setting lock. Push the [Direction] key. Select “513.UL.”. Input a setting value. 「513.UL. 0」 :Modifiable Inputting of the setting value 「513.UL. 1」 :Unable to Read/Write 「513.UL. 2」 :Unable to Write Set the external input/output setting lock.
8 Controlling and adjustment functions FZ-Ex series operetion manual Controlling and adjustment functions Perform setting of the scale ID, the span adjustment and the date and time. Hierarchy of controlling and adjustment functions ” “ can not be used. -58-...
FZ-Ex series operetion manual 8 Controlling and adjustment functions Outputting of the span adjustment result Can not be used. The span adjustment result can be output to a dedicated printer. Select a setting menu. Select the outputting of the span adjustment result. (“2-7-2 Operation of the setting menu, setting of various functions”...
8 Controlling and adjustment functions FZ-Ex series operetion manual Confirm the history. Push the [Enter] key. The indication changes to the “Category of adjustment”. With each pushing of the [Enter] key, the indication changes to “Date”, “Time”, “Temperature”, “Weight used” and “Minimum indication setting”...
FZ-Ex series operetion manual 8 Controlling and adjustment functions Date and time setting Select a setting menu. Select the date and time setting. (“2-7-2 Operation of the setting menu, setting of various functions” and “8-1 Hierarchy of controlling and adjustment functions”) Set the date and time.
8 Controlling and adjustment functions FZ-Ex series operetion manual Password control This function is used for controlling by a password. Select a setting menu. Select the password control setting. (“2-7-2 Operation of the setting menu, setting of various functions” and “8-1 Hierarchy of controlling and adjustment functions”) Set the password control.
FZ-Ex series operetion manual 8 Controlling and adjustment functions 8-11 Password cancellation history This function is used for checking the password cancellation history. 100 history records are stored in all. Select a setting menu. Select the password cancellation history. (“2-7-2 Operation of the setting menu, setting of various functions”...
8 Controlling and adjustment functions FZ-Ex series operetion manual 8-12 Operation of minimum weight indication Can not be used. This is to be set when using the minimum weighed value indication function. Select a setting menu. Select the operation of minimum weighed value indication.
FZ-Ex series operetion manual 8 Controlling and adjustment functions 8-14 Designation of minimum indication Select a setting menu. Select the designation of minimum indication. (“2-7-2 Operation of the setting menu, setting of various functions” and “8-1 Hierarchy of controlling and adjustment functions”) Designate a minimum indication.
Page 76
8 Controlling and adjustment functions FZ-Ex series operetion manual 【Minimum indication list by model】 Setting value 3200 6200 61D .DA. 1 0.001 g 0.01 g 0.01 g 0.1 g 0.1 g SEL .DA. 1 61D .DA. 2 0.002 g 0.02 g 0.02 g 0.2 g 0.2 g...
FZ-Ex series operetion manual 8 Controlling and adjustment functions 8-15 Reset to the factory settings Select a setting menu. Select the factory settings. (“2-7-2 Operation of the setting menu, setting of various functions” and “8-1 Hierarchy of controlling and adjustment functions”) Reset to the factory settings.
Page 78
8 Controlling and adjustment functions FZ-Ex series operetion manual After an indication of “CALWE IT” Select a weight used for the span adjustment. ① appears for one second, the indication changes to the indication of weight selection used for the span adjustment. Push the [Direction] key and select a weight used for the span adjustment.
Page 79
FZ-Ex series operetion manual 8 Controlling and adjustment functions The span adjustment starts. After completion of the zero-point adjustment and the indication changing to “on F.S.”, place the weight in the center of the weighing pan. The indication changes to “PUSH F”. Push the [Function F] key.
8 Controlling and adjustment functions FZ-Ex series operetion manual (1) List of weights used for the span adjustment by model Model FZ623Ex FZ3202Ex FZ6202Ex FZ15001Ex FZ30K0.1GEx name 620 g 3200 g 6200 g 15000 g 30000 g 600 g 3000 g...
FZ-Ex series operetion manual Execution menu 9 Execution menu Operation of the execution menu To operate the execution menu from the state of weighing, chiefly execute the following procedure. (1) Push the [Shift] and [Function F] keys to enter the execution menu from the state of weighing.
Execution menu FZ- Ex series operetion manual Calling of the registered user information This is a function to call the setting that was registered in “9-3 Registration of user information”. Select the execution menu mode. Select the user information calling. (“9-1 Operation of the execution menu”) Select a user to be called.
FZ-Ex series operetion manual Execution menu Registration of user information This is a function to register the content set in “5 User information setting”. Select the execution menu mode. Select the user information registration. (“9-1 Operation of the execution menu”) Select user information to be registered.
Execution menu FZ- Ex series operetion manual Calling of device setting information This is a function to call from the backup memory the setting value of a setting menu item. Select the execution menu mode. Select the calling of device setting information. (“9-1 Operation of the execution menu”) Select the device setting information.
FZ-Ex series operetion manual Execution menu Printing of the GLP header This is a function to add the GLP header at the time of printing. Set the output operation to “412. 00. 1” in the “6-8 Power supply box communication setting”.
Execution menu FZ- Ex series operetion manual Program number and check sum indication Select the execution menu mode. Select the program number and check sum indication. (“9-1 Operation of the execution menu”) Indicate program number. Push the [Direction] key. Select “7.PROG.NO”. Push the [Enter] key.
FZ-Ex series operetion manual Execution menu Outputting of weight data Output to Power supply box: Set the output operation to “412. OO. 1” in “6-8 Power supply box communication setting”. 9-9-1 Outputting of tare weight Select the execution menu mode. Select the output of tare weight.
Execution menu FZ- Ex series operetion manual 9-9-3 Outputting of accumulated value. Output an accumulated value Push the [Transfer] key when accumulated value is indicated in the main LCD. 9-10 Indication of minimum weighed value Can not be used. The minimum weighed value set in the “8-13 Minimum weight indication value setting” can be checked.
FZ- Ex series operetion manual Troubleshooting 10 Troubleshooting 10-1 Error messages Message Cause Countermeasures ・ The weight of an object to be weighed is o-Err ・ Remove the object to be weighed, in excess of the weight of the weighing divide it into two or more, and then capacity.
Page 90
Troubleshooting FZ- Ex series operetion manual Message Cause Countermeasures ・ Sample weight is too light in comparison L-Err with the memorized reference mass of the percent scale. ・ Addition computation was executed ・ Return the indication to zero, t-Err confirm that an asterisk「 」mark doubly due to erroneous adding operation.
Page 91
FZ- Ex series operetion manual Troubleshooting Message Cause Countermeasures ・ User password input is in the wrong. ・ Check the password and input a Err702 correct password. ・ The operation key was pushed at the ・ Do not push the operation key Err703 time of starting from the standby status.
Troubleshooting FZ- Ex series operetion manual 10-2 Troubleshooting Symptom Cause Countermeasures ・ ・ Nothing indicated in DC power supply cable not Check DC power spite of turning on connected supply cable power connection. ・ ・ Indication flashes Power supply box is not switched Make sure that power Erroneous weight on .
Appendix FZ-Ex series operation manual Appendix 1-2 Functional specification Weighing system Tuning fork vibration type Protection class IP65 Type of scale Weight scale / Percent scale / Coefficient scale Functions Adding functions (addition accumulating, net addition, plus side addition, minus side addition) Comparator function (2-point setting, 3-point discrimination, absolute value / relative value discrimination) Buzzer setting, Direct start,...
FZ-Ex series operation manual Appendix Temperature: Operating temperature Scale section and Indicator section : +5 C/ +40 C / humidity Power Supply Box: 0C to +40 C Humidity: 85%RH or lower (with no condensation) Altitude Not higher than 2000 m above sea level Overvoltage category Pollution degree Indicator and scale : 3...
Appendix FZ-Ex series operation manual Appendix 3 Setting menu hierarchy list ■ Hierarchy of functions related to the operation ” “ can not be used. -88-...
Page 99
FZ-Ex series operation manual Appendix ■ Hierarchy of functions related to the performance ■ Hierarchy of user information setting ” “ can not be used. -89-...
Page 100
Appendix FZ-Ex series operation manual ■ Hierarchy of the external input/output functions ” “ can not be used. -90-...
Page 101
FZ-Ex series operation manual Appendix ■ Hierarchy of functions related to the lock -91-...
Page 102
Appendix FZ-Ex series operation manual ■ Hierarchy of controlling and adjustment functions ” “ can not be used. -92-...
FZ-Ex series operation manual Appendix Appendix 4 Print sample ■Span ajustment result English Japanese -93-...
Page 104
Appendix FZ-Ex series operation manual ■ Header English Japanese “MA” is “Minimum weight indication value setting”, See Capter 8 Controlling and adjustment functions. When you have not set up “Minimum weight indication value”, it is printed as “none”. ■ Footer English Japanese -94-...
FZ-Ex series operetion manual Index Index of Terms Actual value setting method ..35, 38, 42 Minus side addition........21 Adding function ......... 21 Addition accumulated ....... 21 Numeric value setting method . 35, 36, 38, 39 Auto power-off ........... 28 Password ..........
Need help?
Do you have a question about the FZ623Ex and is the answer not in the manual?
Questions and answers