DIGITAL MULTIMETER
OPERATOR'S MANUAL
1. Overview
The multimeter is characterized at slim size, portable, stable performance and
anti-dropping capacity. Using 4000 counts digit LCD monitor with character
25mm high, they offer clear readings. With overall circuitry design centering on
large-scale IC A/D converters in conjunction and over-load protection circuit, the
meters give excellent performance and exquisite making as a handy utility
instrument.
The meters can be used to measure DC & AC voltage, DC & AC current,
resistance, capacitor, frequency, duty cycle, Non Contact AC Voltage (NCV)
detection, positive diode voltage fall and audible continuity.
2. Panel Layout
1. Test lead fixture: Fix the test lead.
2. CDS sensor: The CDS sensor can reaction to the ambient brightness range,
then automatically control the LCD backlight to lighten or go out.
3. NCV detection area: Non Contact AC Voltage (NCV) detection area.
4. NCV red light: Non Contact AC Voltage (NCV) detection red light.
5. NCV green light: Non Contact AC Voltage (NCV) detection green light.
6. LCD display: 4000 counts digit, full function symbol display.
7. SELECT key: This key work on the "
diode or continuity test, on the current range, change to DC or AC; If press and
hold SELECT key to power on, "Auto Power Off " function will be disabled.
8. NCV Key: Press and hold the "NCV" key, the meter enters Non Contact AC
Voltage (NCV) detection, the NCV green LED light will light up, free it to exit.
9. HOLD key: Press the "HOLD" key to lock display value, and the "DH" sign will
appear on the display, press it again to exit.
10. REL Key: Press the "REL" key, the meter enters relative measuring mode,
"REL" is displayed on the LCD and the present reading becomes the reference
value and displayed on the display. Relative measurement REL△ =measurement
value-Reference value. Press it again to exit.
11. Hz/% Key: On "ACV/ACA" or "Hz" range, press the "Hz/%" key, you can
choose the Frequency or Duty Cycle measurement.
12. Rotary Switch: Use this switch to select functions and ranges.
13. COM: COM Input Jack
14. 10A: 10A Input Jack
15. VΩmA: V/μAmA/Hz/
16. Crust of meter
17.Protective casing
3. Safety Information
3-1 The meter is designed according to IEC-1010 concerning electronic
measuring instruments with an over-voltage category 600V (CATⅢ) and
pollution 2.
3-2 Follow all safety and operating instructions to ensure that the meter is used
safely and is kept in good operating condition.
3-3 safety symbols:
Important safety information, refer to the operating manual.
Dangerous voltage may be presence.
Double insulation (protection Class II)
4. Special Cautions for Operation
4-1 The meters can be safe only according to standard procedures when used in
conjunctions with the supplied test leads. To replace damaged test leads with
only the same model or same electric specifications.
4-2 To avid risk of electric shock, do not use the meters before the cover is in
place.
4-3 The range switch should be right position for the testing.
4-4 To avoid electric shock and damaging the instruments, the input signals are
forbidden to exceed the specified limits.
4-5 When measuring TV set or switched power, attention should be paid to the
possible pulses that may bring destruction to the circuit.
4-6 Range switch position is forbidden to be changed at random during
measurement.
4-7 Take caution against shock in the course of measuring voltage higher than
DC 60V & AC 30V.
4-8 Protection fuse should be replaced only with same type and same
specification.
4-9 After operation is finished, set function switch at OFF to save battery power.
39K
" range, press the key to choose
Input Jack
4-10 If the meter is without usage for long time, take out battery to avoid damage
by battery leakage.
5. GENERAL SPECIFICATIONS
5-1 Max Voltage between input terminal and Earth Ground: CAT Ⅲ 600V
5-2 Over-range Indication: display "OL" for the significant digit.
5-3 Automatic display of negative polarity "-" .
5-4 Low Battery Indication: "
" displayed.
5-5 Max LCD display: 4000 counts digit.
5-6 Auto range control
5-7 Auto Power Off: When measurement exceeds 15 minutes without switching
mode and pressing key, the meter will switch to standby mode. Press any
key to exit standby mode. When restart the system, press and hold SELECT
key to disable auto power off.
5-8 Auto LCD backlight
5-9 Fuse protection: 500mA/500V Fast Fuse, 10A/500V Fast Fuse
5-10 Power supply: 1.5V×2 "AA" R6P battery
5-11 Operating Temp.: 0℃ to 40℃ (relative humidity <85%)
5-12 Storage Temp.: -10 ℃ to 50℃ (relative humidity <85%)
5-13 Guaranteed precision Temp.: 23±5℃ (relative humidity <70%)
5-14 Dimension: 150x100x36mm
5-15 Weight: approx. 250g (including battery)
6. Testing Specifications
Accuracy is specified for a period of year after calibration and at 18 ℃ to 28 ℃
(64℉ to 82℉) with relative humidity to 70%.
6-1 DC Voltage
Range
Resolution
400mV
0.1mV
4V
1mV
40V
10mV
400V
100mV
600V
1V
-- Impedance: 10MΩ, More than 100MΩ on 400mV range
-- Overload protection: 600V DC or AC rms
6-2 AC Voltage
Range
Resolution
4V
1mV
40V
10mV
400V
100mV
600V
1V
-- Impedance: 10MΩ
-- Overload protection: 600V DC or AC rms
-- Frequency Range: 40 to 400Hz
-- Response: average, calibrated in rms of sine wave
6-3 DC Current
Range
Resolution
400μA
0.1μA
4000μA
1μA
40mA
10μA
400mA
100μA
4A
1mA
10A
10mA
-- Overload protection: 500mA/500V Fast Fuse
10A/500V Fast Fuse, 10A up to 10 seconds
6-4 AC Current
Range
Resolution
400μA
0.1μA
4000μA
1μA
40mA
10μA
400mA
100μA
4A
1mA
10A
10mA
-- Overload protection: 500mA/500V Fast Fuse
10A/500V Fast Fuse, 10A up to 10 seconds
-- Frequency Range: 40 to 400Hz
-- Response: average, calibrated in rms of sine wave
6-5 Resistance
Range
Resolution
400Ω
0.1Ω
4kΩ
40kΩ
10Ω
400kΩ
100Ω
4MΩ
1kΩ
40MΩ
10kΩ
-- Overload protection: 500V DC or AC rms
6-6 Capacitance
Range
Accuracy
5.12nF
±(3.0% of rdg + 10 digits)
51.2nF
512nF
±(2.5% of rdg + 5 digits)
5.12µF
51.2µF
200µF
±(5.0% of rdg + 10 digits)
-- Overload protection: 500V DC or AC rms
Accuracy
±(0.5% of rdg + 2 digits)
±(0.8% of rdg + 2 digits)
Accuracy
±(0.8% of rdg + 3 digits)
±(1.2% of rdg + 3 digits)
Accuracy
±(1.2% of rdg + 2 digits)
±(2.0% of rdg + 3 digits)
Accuracy
±(1.5% of rdg + 3 digits)
±(2.5% of rdg + 5 digits)
Accuracy
±(1.0% of rdg + 3 digits)
±(1.0% of rdg + 2 digits)
±(1.5% of rdg + 3 digits)
Resolution
1pF
10pF
100pF
1nF
10nF
100nF

Advertisement

loading

Summary of Contents for HoldPeak 39K

  • Page 1 4-10 If the meter is without usage for long time, take out battery to avoid damage by battery leakage. DIGITAL MULTIMETER 5. GENERAL SPECIFICATIONS OPERATOR’S MANUAL 5-1 Max Voltage between input terminal and Earth Ground: CAT Ⅲ 600V 1. Overview 5-2 Over-range Indication: display “OL”...
  • Page 2 2. When the input is not connected, i.e. at open circuit, the figure ‘OL’ will be 6-7 Frequency displayed for the over-range condition. Range Accuracy Resolution 3. When checking in-circuit resistance, be sure the circuit under test has all 5.12Hz 0.001Hz power removed and that all capacitors have been discharged fully.