The following are the user preventative maintenance tasks. It is recommended that these be
performed on a regular basis at a frequency determined by the usage of the equipment, but not less
than once every month.
WARNING
:
Before undertaking any of these tasks disconnect the unit from the mains.

General

Check all cables, connectors and transducers for damage and repair or
replace where necessary. The repair may involve your local service centre,
supplier or Ultrasound Technologies Ltd. For advise on any damaged part
contact them immediately.
Cleaning - Enclosure
Clean the exterior of the system with a soft dry cloth. In the event of
stubborn spots, disconnect the system from the power source. Use a soft
cloth that has been dampened - not soaked - in a mild detergent solution or
isopropyl alcohol. Be sure to keep excess moisture from entering the
cabinet via any openings that may be present.
Cleaning - Transducers
Use a cloth dampened in a mild detergent solution or isopropyl alcohol to
clean the transducer and cable. Remove all traces of the detergent or
alcohol by wiping with a cloth dampened in clear water. Never soak the
transducer cable or connector.
WARNING:
Transducers must never be exposed to gas or heat sterilization or be left
immersed in any liquid for more that a few seconds.
Fetatrack310 Service Manual Issue3
©Ultrasound Technologies Ltd, Lodge Way, Portskewett, Caldicot, South Wales NP26 5PS, United Kingdom.
T +44 (0) 1291 425425
Ultrasound Technologies Ltd
F: +44 (0) 1291 427093

Maintenance

E: ultratec@doppler.co.uk
Page 14
W:
www.doppler.co.uk