Troubleshooting; No Display Information On Lcd; Keyboard Does Not Respond; No Sound From Loudspeakers - Ultrasound Technologies Fetatrack 310 Service Manual

Doppler fetal monitor; issue 3
Hide thumbs Also See for Fetatrack 310:
Table of Contents

Advertisement

The information in this section will help you to check and correct common operation and system
problems. Refer to the troubleshooting hints which deal with your problem. Perform the suggested
steps. If the problem is not solved, check once again to make sure that you have used all of the
suggested steps to resolve the problem.
Electronic failures and service procedures are not included in this manual, as all servicing of the
system must be performed by a qualified service technician. Valuable time however can be saved
by documenting the problem .
In general, when you have a problem, check your control settings to be sure that they are in proper
operating position.
Consult the appropriate section in this manual for specific information on
particular controls or operating modes.
WARNING:
Disconnect system from the power source before checking fuses and connections.
Check all connections and fuses. Replace fuses with same type and rating as indicated on the rear
panel of the unit.

No display information on LCD

·
Verify the system is on and that the fuses are intact.:

Keyboard does not respond

·
Reset system by turning off then back on.
·
Verify the system is on and that the fuses are intact.
No sound from loudspeaker.
·
Verify the system is on and that the fuses are intact.
·
Check volume control is set high.
·
Check that the correct transducer has been selected.
·
Check connection of the transducer.
No FHR information on display or FH trace printed on recorder.
·
Check that the correct transducer has been selected and connected to the
patient.
·
Check connection of the transducer.
·
Check for audio FH complex and reposition transducer until clearly heard.

No chart printed .

·
Verify the system is on and that the fuses are intact.
·
Check that the recorder is turned on.
·
Check paper is inserted into the recorder correctly.
·
Check recorder door is correctly closed
·
Check that paper moves out of the recorder at the correct rate.
Fetatrack310 Service Manual Issue3
©Ultrasound Technologies Ltd, Lodge Way, Portskewett, Caldicot, South Wales NP26 5PS, United Kingdom.
T +44 (0) 1291 425425
Ultrasound Technologies Ltd
F: +44 (0) 1291 427093

Troubleshooting

E: ultratec@doppler.co.uk
Page 13
W:
www.doppler.co.uk