Observed authority requirements
EMC 2004/108/EC ............. EN 61326-1
LVD 2006/95/EC ................ EN 61010-1
UL, Standard for Safety ...... UL 508
EAC TR-CU 020/2011 ........ EN 61326-1
of span = of the currently selected measurement range
Display readout on the DRSL-DISPLAY of sensor error
detection and input signal outside range
Device:
OUT.ERR=NONE.
DRST-UN
Outside range readout (IN.LO, IN.HI):
If the valid range of the A/D converter or the polynomial is exceeded
Input
0...1 V / 0.2...1 V
VOLT
0...10 V / 2...10 V
CURR
0...20 mA / 4...20 mA
0...800 Ω
LIN.R
0...10 kΩ
POTM
TEMP
TC / RTD
Display readout below min.- / above max. (-1999, 9999):
Input
All
Standard
Sensor error check:
Configuration
Else:
Range
Readout
IN.LO
IN.LO
IN.LO
IN.LO
IN.LO
IN.LO
-
IN.LO
Range
All
17
Sensor error detection:
OFF
ON
Limit
< -25 mV
IN.HI
> 1.2 V
< -25 mV
IN.HI
> 12 V
< -1.05 mA
IN.HI
> 25.05 mA
< 0 Ω
IN.HI
> 1075 Ω
< 0 Ω
IN.HI
< 110 kΩ
< -0.5 %
IN.HI
> 100.5 %
< temperature range -2°C
IN.HI
> temperature range +2°C
Readout
Display readout
-1999
<-1999
Display readout
9999
8
Specifications
Limit
>9999