Table of Contents

Advertisement

Available languages

Available languages

Weight
Operating
temperature
Operating
relative
humidity
Hematocrit range
Data port

Limitations

The 2in1. MaX meter, test strips and control solution are been
designed, tested and proven to work together effectively to
provide accurate blood glucose measurements. Do not use
components from other brands.
• Fresh capillary blood may be collected into test tubes
containing sodium heparin, lithium heparin if the blood is
used within 10 minutes. Do not use sodium fluoride/oxalate
or other anticoagulants or preservatives.
• Use only with whole blood. Do not use with serum or plasma
samples.
• Very high (above 70%) hematocrit levels can cause false
results. Talk to your health care professional to find out your
hematocrit level.
• The system is tested to accurately read the measurement of
glucose in whole blood within the range of 10 to 600 mg/dL
(0.6-33.3 mmol/L).
• Fatty substances (triglycerides up to 3,000 mg/dL (166.7
mmol/L) or cholesterol up to 500 mg/dL (27.7 mmol/L))
have no major effect on blood glucose test results.
• The 2in1. MaX Blood Glucose Monitoring System has been
tested and shown to work properly up to 10,000ft (3,048
meters).
Approximately 47g (with battery
installed)
5-45°C
10-90% (non-condensing)
0-70%
Micro USB
45

Hide quick links:

Advertisement

Table of Contents
loading

Related Products for 2in1 Max

Table of Contents