THOMAS SPLINT CLEANING INSTRUCTIONS
Cleaning can be achieved by either manual or automated methods - manual cleaning should only be undertaken
when automated methods are inappropriate or unavailable. Automated cleaning is a controlled process that will
allow more consistent results. Where available the use of mechanical cleaners such as washer-disinfectors and
ultrasonic tanks is preferred to the manual cleaning of items. The cleaning method should be assessed to ensure
the effectiveness of the cleaning process without damage to equipment.
Medical Devices Agency guidelines on the method of decontamination taking into account infection risk from the
reprocessed device to the patient
Risk
Application of an item
High
In close contact with a break in the skin or mucous membrane. Introduced into
sterile body areas.
Medium
In contact with mucous membranes. Contaminated with particularly virulent or
readily transmissable organisms prior to use on immunocompromised patients.
Low
In contact with healthy skin. Not in contact with patient.
CLEANING GUIDELINES
Please refer to guidelines below for washing instructions.
Thomas Splint Component
Recommended cleaning method
Thomas Splint side members
Cleaning & Steam Sterilisation
Thomas Splint Thigh Hoops
Thermal Washer Disinfector
Thomas Splint Diposable Slings and Hoop Covers
Single Patient Use 400 wash
MECHANICAL CLEANING -THERMAL WASHER DISINFECTORS (REPEATED EXPOSURE TO WET HEAT)
Initial Clean - at or below 35°C. Followed by a hot water disinfection rinse where the surface temperature of the
item processed should reach a minimum temperature of 71°C for a minimum of 3 minutes, 80°C for 1 minute or
90°C for 1 second (HC(91)33 and BS 2745)
MANUAL CLEANING - NON IMMERSION
A warm water/detergent solution at correct dilution
A clean, disposable, absorbent, non-shedding cloth or mechanical drying facility (e.g. drying cabinet or industrial
hot air dryer)
A chemical neutraliser, first aid kit and eye wash bottle, in case of splashing with detergent.
STEAM STERILISATION
Direct contact between the material being sterilised and
Sterilisation Temperature (°C)
pure dry saturated steam at the required temperature in
Max Allowance Temperature (°C)
the absence of air. The recommended combinations of
Minimum Holding Time (minutes)
time and temperature are given in HTM 2010, Part 3:
Frame Measurements for Steam Sterilisation
(with side members fully retracted)
Adult
Paediatric
NOTE: The Thomas Splint System is designed for long
term use, however regular inspection of adjustment areas
and frame body is advised. Please replace if necessary.
THOMAS SPLINT SLINGS
All Thomas Splint slings are single patient use and must be discarded
at the end of the treatment.
Össur Americas
27051 Towne Centre Drive
Foothill Ranch, CA 92610, USA
Tel: +1 (949) 382 3883
Tel: +1 800 233 6263
Fax: +1 800 831 3160
ossurusa@ossur.com
Össur Canada
2150 – 6900 Graybar Road
Recommendation
Richmond, BC
Cleaning and Sterilisation
V6W OA5 , Canada
Tel: +1 604 241 8152
Fax: +1 866 441 3880
Sterilisation or disinfection required. Cleaning
may be acceptable in some agreed situations
Össur Europe BV
Cleaning (physically) removes infectious
De Schakel 70
agents without necessarily destroying them.
5651 GH Eindhoven
The Netherlands
Tel: +800 3539 3668
Tel: +31 499 462840
Fax: +31 499 462841
info-europe@ossur.com
Össur Deutschland GmbH
Augustinusstrasse 11A
50226 Frechen
Deutschland
Tel: +49 (0) 2234 6039 102
Fax. +49 (0) 2234 6039 101
info-deutschland@ossur.com
115
121
126
134
118
124
129
137
30
15
10
3
Length
Width
80
21 cm
56
19cm
Össur hf.
Grjótháls 1-5
110 Reykjavík
40C
Iceland
Wash 40°C
Do Not Blech
Dry Flat, Do Not Wring
FOR SINGLE
PATIENT USE
Do Not Iron
Single Patient Use Only
WWW.OSSUR.COM
Össur Nordic
Össur Europe BV – Italy
P.O. Box 67
Via Baroaldi, 29
751 03 Uppsala, Sweden
40054 Budrio, Italy
Tel: +46 1818 2200
Tel: +39 05169 20852
Fax: +46 1818 2218
Fax: +39 05169 22977
info@ossur.com
orders.italy@ossur.com
Össur Iberia S.L.U
Össur APAC
Calle Caléndula, 93 -
2F, W16 B
Miniparc III
No. 1801 Hongmei Road
Edificio E, Despacho M18
200233, Shanghai, China
28109 El Soto de la Moraleja,
Tel: +86 21 6127 1707
Alcobendas
Fax: +86 21 6127 1799
Madrid – España
asia@ossur.com
Tel: 00 800 3539 3668
Fax: 00 800 3539 3299
Össur Australia
orders.spain@ossur.com
26 Ross Street,
orders.portugal@ossur.com
North Parramatta
NSW 2151 Australia
Össur UK Ltd
Tel: +61 2 88382800
Unit No 1
Fax: +61 2 96305310
S:Park
infosydney@ossur.com
Hamilton Road
Stockport SK1 2AE, UK
Tel: +44 (0) 8450 065 065
Fax: +44 (0)161 475 6321
ossuruk@ossur.com
©Copyright Össur 2017
IFU 1322_001
Rev. 1
Instructions for Use
THOMAS SPLINT

Advertisement

loading
Need help?

Need help?

Do you have a question about the THOMAS SPLINT and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Ossur THOMAS SPLINT

  • Page 1 THOMAS SPLINT SLINGS Wash 40°C Do Not Blech Dry Flat, Do Not Wring FOR SINGLE All Thomas Splint slings are single patient use and must be discarded PATIENT USE at the end of the treatment. Do Not Iron Single Patient Use Only WWW.OSSUR.COM...
  • Page 2 FITTING INSTRUCTIONS Easily adjustable to a number of leg lengths, the Pearson attachment can be added to the Thomas Splint to offer Selecting the correct sized splint and hoop is made easier as the system is designed to be adjustable. The splint earlier mobilisation for your patient.