Badger Meter FC-5000 User Manual page 33

Btu monitor, industrial flow computer
Hide thumbs Also See for FC-5000:
Table of Contents

Advertisement

3� Press UP/DOWN to highlight a temperature sensor calibration option, then
press ENTEO�
The calibration setting options are described below� The settings that appear on
the device are relative to the sensor type and will only show if a sensor type
is chosen�
Option
ALPHA COEFF
BETA COEFF
DELTA COEFF
OHMS: LOW
TEMP: LOW
OHMS: HIGH
TEMP: HIGH
OFFSET
COEFF A
COEFF B
COEFF C
N
OTEE:
The temperature sensor calibration for Temperature Sensor 2 includes an
adjustment field "TEMP 2 ADJ (° F)"� Changing this setting will increase or
decrease the temperature variance by the value entered, in
degrees Fahrenheit�
February 2018
Description
Callendar-Van Dusen coefficients
Resistance (Ω) at 0° C / Resistance (Ω) at "TEMP: LOW"
Temperature at "OHMS: LOW" (° C)
Resistance (Ω) at temperature "TEMP: HIGH"
Temperature at "OHMS: HIGH" (° C)
Temperature calibration offset
Steinhart-Hart coefficients
Table 21: Temperature sensor calibration descriptions
CTL-UM-02038-EN-02
Advanced Setup
Page 33