Q Imaging EXi Aqua User Manual page 13

Bio-imaging microscopy camera for highly versatile, high performance imaging
Table of Contents

Advertisement

Connecting external sources to the camera
Before you can begin to externally trigger your camera, you need to connect the necessary trigger signal
source to the camera. To do this you need to ensure that you have a 6-pin miniature circular DIN cable and
ensure that the pin-outs are set correctly from your trigger source. Refer to Appendix A for further details on
pin assignments and a diagram of the mini-DIN connector.
There are several ways that you can provide an external pulse to the camera, one of them being via the
QImaging External Trigger Board, which is available through QImaging, and another common device is a
function generator. To trigger the camera, you need to provide a 5V TTL pulse though the pin stated in
Appendix A.
To ensure that you will use the external triggering mechanism correctly, it is important to ensure that the
following questions have been answered when setting up your system design:
• Has a system design been developed that incorporates the necessary elements such as:
a timing diagram;
a simple state diagram or "flow of events";
necessary components including microscope, flash lamp, shutter, etc.;
a software solution for extracting and/or processing your images; and
and external triggering source (e.g. QImaging trigger board, function generator, etc.).
• Have you determined the type of triggering mode you require for your application?
• Are you sure the settings in the camera have been set correctly? For instance, are you monitoring
the SYNC B output, and if so have you set the correct output mode for your system?
The above list should be used as a guide and should be consulted before contacting QImaging Technical
Support.
Trigger Mode
Continual
Internal to Camera
Edge Hi/Lo
External Trigger Port
Pulse Hi/Lo
External Trigger Port
Strobe Hi/Lo
External Trigger Port
Software
QCAM API Trigger Event
Table 1—Trigger Sources and Integration Modes
Trigger Source
Exposure Control
Internal to Camera
Internal to Camera
Length of User Pulse
Length of User Pulse
Internal to Camera
9