Technical Data - TESTO 112 Instruction Manual

Ntc-/pt100 measuring instrument
Hide thumbs Also See for 112:
Table of Contents

Advertisement

16

9. Technical data

9.
Instrument
Characteristic
Parameters
Measuring range
Resolution
Accuracy
Probe
Measuring rate
Operating temperature range -20...+50°C / -4...+122°F
Storage temperature
Voltage supply
Battery life
Protection class
EC Directive
Warranty
System accuracy
Measuring range
Measuring instrument + NTC temperature probe
-50.0°C...-25.1°C
-25.0...+40.0°C
+40.1...+80.00°C
+80.1...+120.0°C
Measuring instrument + Pt100 temperature probe
-50.0...-25.1°C
-25.0...+40.0°C
+40.1...+140.0°C
+140.1...+200.0°C
+200.1...+300.0°C
roducts Online at:
Technical data
Value
Temperature (°C/°F)
Pt100 probe: -50...+300°C / -58...+572°F
NTC probe: -50...+120°C / -58...+248°F
0.1°C / 0.1°F
1x mini DIN socket for Pt100 or NTC temperature probe
2/s
-30...+70°C / -22...+158°F
1x 9V monobloc battery/rech. battery
approx. 70h
with TopSafe (accessory part) and probe connected: IP65
89/336/EEC
2 years
Instrument
±1% of reading
±0.2°C
±0.3°C
±0.5°C
±0.2°C
±0.2°C
±0.2°C
±0.2°C
±0.3°C
GlobalTestSupply
www.
S
See
YSTEM ACCURACY
Probe
±0.7% of reading ±1.8% of reading
±0.2°C
±0.4°C
±0.6°C
±0.3°C
±0.2°C
±0.4°C
±0.6°C
±0.8°C
System
±0.5°C
±0.8°C
±1.2°C
±0,6°C
±0.5°C
±0.7°C
±0.9°C
±1.2°C
.com