Chapter 7
actinic illumination without preceding saturation pulse analysis
(Initial pulse). Width of actinic illumination is adjusted as de-
scribed above for clock interval and initial pulse is selected by
the SET command. Further, actinic light intensity is adjusted in
the window "Light Sources" (Fig. 41, page 77).
An Actinic + Yield routine can be started using the START but-
ton on the side panel of Fig. 45. In this case, the screen display
will automatically switch to the Actinic + Yield experimental win-
dow (Fig. 32, page 60).
Actinic + Yield Settings
Width (mm:ss)
Initial Pulse
Fig. 45:
Actinic + Yield Settings
7.3.3.2

Induction Curve

Induction curve experiments are configured in the window "In-
duction Curve Settings" (Fig. 46, page 83).
Delay (range 5 s to 10 min) defines the dark interval between
saturation pulse analysis with the dark-acclimated sample (F
F
determinations, Section 5.1.1, page 32) and beginning of ac-
M
tinic illumination.
82
Touchscreen Operation
START
2:00
yes
SET
EXIT
Side Panel
START. Start Actinic
light + Yield routine and
switch to Actinic Light
and Yield window (Fig.
32, page 60).
,
. Up/Down
keys to move to a menu
line.
SET. Select item.
,
0