Siemens SITRANS LG270 Operating Instructions Manual page 66

Guided wave radarfoundation fieldbus coax probe, minus 196...plus 280 degrees c; minus 196 ... plus 450 degrees c
Table of Contents

Advertisement

Shock resistance
Ʋ Coax probe
Electromechanical data - version IP 66/IP 67 and IP 66/IP 68; 0.2 bar
Options of the cable entry
Ʋ Cable entry
Ʋ Cable gland
Ʋ Blind plug
Ʋ Closing cap
Material ca-
Material seal
ble gland
insert
PA
NBR
Brass, nickel-
NBR
plated
Stainless
NBR
steel
Wire cross-section (spring-loaded terminals)
Ʋ Massive wire, stranded wire
Ʋ Stranded wire with end sleeve
Electromechanical data - version IP 66/IP 68 (1 bar)
Options of the cable entry
Ʋ Cable gland with integrated connec-
tion cable
Ʋ Cable entry
Ʋ Blind plug
Connection cable
Ʋ Wire cross-section
Ʋ Wire resistance
Ʋ Tensile strength
Ʋ Standard length
Ʋ Max. length
Ʋ Min. bending radius
Ʋ Diameter
Ʋ Colour - Non-Ex version
Ʋ Colour - Ex-version
Display and adjustment module
Display element
66
25 g, 6 ms according to EN 60068-2-27 (mechanical
shock) with tube length 50 cm (19.69 in)
M20 x 1.5; ½ NPT
M20 x 1.5; ½ NPT (cable ø see below table)
M20 x 1.5; ½ NPT
½ NPT
4.5 ... 8.5 mm
5 ... 9 mm
0.2 ... 2.5 mm² (AWG 24 ... 14)
0.2 ... 1.5 mm² (AWG 24 ... 16)
M20 x 1.5 (cable: ø 5 ... 9 mm)
½ NPT
M20 x 1.5; ½ NPT
0.5 mm² (AWG 20)
< 0.036 Ω/m
< 1200 N (270 lbf)
5 m (16.4 ft)
180 m (590.6 ft)
25 mm (0.984 in) with 25 °C (77 °F)
approx. 8 mm (0.315 in)
Black
Blue
Display with backlight
SITRANS LG270 - Operating Instructions
Cable diameter
6 ... 12 mm
7 ... 12 mm
10 ... 14 mm
PBD-51041071

Advertisement