Info - Device ID
Info - Sensor character-
istics
Backup on paper
Backup in the display and
adjustment module
42
In this menu item, the identification number of the instrument in a
Foundation Fieldbus system is shown.
In this menu item, the features of the sensor such as approval, pro-
cess fitting, seal, measuring range, electronics, housing and others
are displayed.
Example for displayed sensor features.
6.5

Saving the parameterisation data

We recommended writing down the adjustment data, e.g. in this op-
erating instructions manual, and archiving them afterwards. They are
thus available for multiple use or service purposes.
If the instrument is equipped with a display and adjustment module,
the data in the sensor can be saved in the display and adjustment
module. The procedure is described in menu item "Copy device
settings" in the menu "Additional settings". The data remain there
permanently even if the sensor power supply fails.
The following data or settings for adjustment of the display and ad-
justment module are saved:
All data of the menu "Setup" and "Display"
The items "Sensor-specific units, temperature unit and linearisa-
tion" in the menu "Additional settings".
The values of the user-programmable linearisation curve
The function can also be used to transfer settings from one instru-
ment to another instrument of the same type. If it is necessary to
exchange a sensor, the display and adjustment module is inserted
into the replacement instrument and the data are likewise written into
the sensor via the menu item "Copy device settings".
SITRANS LG270 - Operating Instructions
PBD-51041071