Sibelmed Sibelsound 400 Quick Start Manual
Sibelmed Sibelsound 400 Quick Start Manual

Sibelmed Sibelsound 400 Quick Start Manual

Audiometer
Hide thumbs Also See for Sibelsound 400:

Advertisement

Press F3, choose the
+
+
signal:
continuous (C), pulse (P) and press
Choose the
signal intensity.
Use F1 to choose the
+
+
(AUX) or technician microphone (MIC) and press
1 s
Choose the
number of words
+
+
1 s
volume of the input device
Choose the
auxiliary (AUX) or technician microphone (MIC) and
+
+
press
1 s
Choose the number of
To set the counter to zero, press

COMMON FUNCTIONS

Applying the signal
or
Choose
inverted mode
+
+
the SIGNAL key is not pressed).
(Press simulta-
Threshold memorisation.
Threshold revision,
and the frequency with
Printing
the audiometric report (the thresholds
have to be memorised).
Save a test
to the database.
Accessing the Database menu if there are no
thresholds.
Activating the
the signal from the technician microphone
and vice-versa). Choose the volume for the
technician's
patient'smicrophone. (M.PA)
0318
Product class IIa according to Directive 93/42/EEC
SIBEL S.A., Rosselló 500, 08026 Barcelona (Spain)
Domestic Sales: Tel. 93 436 00 08 - E-mail: comercial@sibelmed.com
International Sales: Tel. +34 93 436 00 07 - E-mail: export@sibelmed.com
Technical Assistance: Tel. +34 93 433 54 50 - E-mail: sat@sibelmed.com
SIBEL S.A. is a member of the
Fax: +34 93 436 16 11
www.sibelgroup.com
A LGO C A LGO C
presentation mode for the
CONTIN.
o
.
A LGO C A LGO C
60
input
device: auxiliary
NWF1:AUX/MICVD
10
o
o
.
NP F1: AUX/MIC VD
(NW) and press
10
o
o
.
(VD):
NWF1:AUX/MIC VD
10
o
o
.
correct
words.
A LGO C A LGO C
60
.
to the patient.
(the signal is enabled when
A HZ C A HZ C
60
TEST
choose the conduction with
TH R ESH O L D S AV ED
A --- XXX A ---
60
TEST
PR IN TIN G
TEST
intercom
(the patient receives
INTERCOM
E.TC 03
earphones
(E.TC)
and
the
www.sibelmed.com
CONTIN.
60
00/10
03
Congratulations for purchasing an audiometer from SIBELMED, a leading company in the sector.
To optimise the use of this device, please refer to the USER'S MANUAL.
03
The SIBELSOUND 400
clinical audiometer is a compact two-channel device, the main components of which are a tone
generator, a noise generator, a set of earphones for air conduction, a vibrator for bone conduction
03
and an alphanumeric liquid crystal screen.
The whole system is controlled by a Digital Signal Processor DSP that allows audiometric
examination to be carried out quickly, simply and reliably to establish hearing thresholds and to
carry out screening tests such as suprathreshold tonal tests.
02/10
60

INSTALLATION

Install the device somewhere:
- With a 90V - 264V mains outlet and ground connection (required power is < 50 VA).
- With room temperature between 10 ºC and 40 ºC.
- With relative humidity lower than 90% without condensation.
- Next to a soundproof cabin or in a room with a very low level of ambient noise (another possibility is
placing noise dampers in the earphones).
- Away from possible splashes of water or any other liquid.
- Free from objects that prevent air flow during operation.

MENUS

1000
60
The audiometer has several options that can be accessed by pressing the
XXXXX
1. TESTS
- 1. TONE
- 2. Sisi*
1000
60
- 3. Speech*
- 4. Fowler*
XXXXX
- 5. Tone Decay* - 5. Sel. Freq.
- 6. Luscher*
- 7. Weber*
XXXXX
- 8. Pure Tone
SAVE D
- 9. Free
M.PA 03
*Depending on models
** Permitted in models,
AOM, AOM+ AND SUPRA
Press
Press
previous menu (press this key for more than 1 second to return to the screen of the selected test).
Quick Start Guide
SIBELSOUND
2. CONFIGURATION
3. MAINTENANCE
- 1. SAVE CONFIG.
- 1. TEST DEVICE
- 2. Modif Default
- 1. CPU
- 3. Reset Default
- 2. LCD
- 4. Diagnosis
- 3. KEYBOARD
- 4. Printer
- 1. Freq. Sel.
- 5. Update key
- 2. Musical Freq.
- 6. Version
*- 6. Printer
- 7. RESET DEVICE
HF
- 7. Language
- 2. CALIBRATION
- 8. CONTRAST
- 1. ANSI / ISO
- 9. DATE - TIME
- 2. CAL. WARNING
- 3. CAL. DEFECT
- 3. MONITOR*
- 4. LOGO-INTER*
- 5. AUXILIARY*
or turn
to move around the various options of the menu.
or
to access the selected option and press
400
key, as follows:
4. DATABASE*
- 1. Display
- 2. Delete DB
- 3. Find patient
- 4. Delete Patient.
5. REF TONE**
6. AUTO MASK
to return to the

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Sibelsound 400 and is the answer not in the manual?

Questions and answers

Summary of Contents for Sibelmed Sibelsound 400

  • Page 1 Use F1 to choose the input device: auxiliary NWF1:AUX/MICVD (AUX) or technician microphone (MIC) and press Congratulations for purchasing an audiometer from SIBELMED, a leading company in the sector. NP F1: AUX/MIC VD Choose the number of words (NW) and press To optimise the use of this device, please refer to the USER'S MANUAL.
  • Page 2 PERFORMING AN AUDIOMETRIC TEST, step by step Press F3 and choose the presentation mode for AHZ C the signal: continuous (C) or pulse (P) and press CONTIN. CONTIN. 1. Make sure that patient sitting comfortably, relaxed rested paying as much attention as possible to the test. signal intensity.