Download Print this page
Milwaukee MI 404 User Manual
Milwaukee MI 404 User Manual

Milwaukee MI 404 User Manual

Free & total chlorine and ph meter
Hide thumbs Also See for MI 404:

Advertisement

Quick Links

USER MANUAL
MI 404
Free & Total Chlorine
and pH Meter
Dear Customer,
Thank you for choosing a Martini product.
This manual will provide you with the
necessary information for the correct use of
the nstrument. Please read it carefully
before using the meter. This instrument is in
compliance with
directives.
SPECIFICATIONS:
Range
0.00 to 5.00 mg/L Cl2Cl2
Resolution
0.01 mg/L (0.00-3.50 mg/L);
0.10 mg/L (above 3.50 mg/L)
Precision
±0.04 mg/L @ 1.50 mg/L
Light Source
Tungsten lamp
Light Detector Silicon Photocell and 525
nm narrow band interfere-
nce filter
Method
Adaptation of the USEPA
method 330.5 and Standard
Method 4500-Cl G.
Environment
0 to 50°C (32 to 122°F);
100% RH max.
Battery Type
1 x 9 volt
Auto-Shut off
After 10' of non-use
Dimensions
192 x 104 x 52 mm
(7.5 x 4.1 x 2")
Weight
380 g
MEASUREMENT PROCEDURE:
1• Turn the meter on by pressing ON/OFF.
2• When the LCD displays
"- - -", the meter is ready.
3• Fill the cuvet with 10 mL
of unreacted sample, up to
the mark, and replace the
cap.
4• Place the cuvet into the
holder and ensure that the
notch on the cap is
positioned securely into the
groove.
5• Press ZERO and "SIP" will
blink on the display.
6• After a few seconds the display
will show "-0.0-". The meter is
now zeroed and ready for measurement.
7• Remove the cuvet.
8•
For free chlorine measurement: add 3
drops of DPD 1 indicator and 3 drops of
DPD 2 buffer to an empty cuvet.
Immediately add 10 mL of unreacted sample.
Replace the cap and swirl gently.
9• Immediately reinsert the cuvet into the
holder and ensure that the notch on the
cap is positioned securely into the
groove.
10• Press READ and "SIP" will blink during
measurement.
11• The instrument directly displays
concentration
in mg/L of free chlorine.
12• Then, for total chlorine measurement,
remove the cuvet and add to the
reacted sample 3 drops of DPD 3.
Replace the cap and swirl gently.
13• Reinsert the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
14• Wait for 2 minutes and 30 seconds then
press READ and "SIP" will blink during
measurement.
15• The instrument directly displays
concentration in mg/L of total chlorine.
Interferences:
• Bromine (positive error)
• Chlorine dioxide (positive error)

Advertisement

loading
Need help?

Need help?

Do you have a question about the MI 404 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Milwaukee MI 404

  • Page 1 2• When the LCD displays groove. MI 404 the nstrument. Please read it carefully “- - -”, the meter is ready. 10• Press READ and “SIP” will blink during before using the meter.
  • Page 2 • Iodine (positive error) On sample reading Detach the battery from “-bA-”, the battery is dead and the terminals and • Oxidized Manganese (positive error) must be replaced. Once • “-SA-”, there is too much attach a fresh 9V • Chromium (positive error) this indication is displayed, light for the sample battery while paying...