MI 404
Free & Total Chlorine
and pH Meter
Dear Customer,
Thank you for choosing a Martini product.
This manual will provide you with the
necessary information for the correct use of
the nstrument. Please read it carefully
before using the meter. This instrument is in
compliance with
directives.
SPECIFICATIONS:
Range
0.00 to 5.00 mg/L Cl2Cl2
Resolution
0.01 mg/L (0.00-3.50 mg/L);
0.10 mg/L (above 3.50 mg/L)
Precision
±0.04 mg/L @ 1.50 mg/L
Light Source
Tungsten lamp
Light Detector Silicon Photocell and 525
nm narrow band interfere-
nce filter
Method
Adaptation of the USEPA
method 330.5 and Standard
Method 4500-Cl G.
Environment
0 to 50°C (32 to 122°F);
100% RH max.
Battery Type
1 x 9 volt
Auto-Shut off
After 10' of non-use
Dimensions
192 x 104 x 52 mm
(7.5 x 4.1 x 2")
Weight
380 g
MEASUREMENT PROCEDURE:
1• Turn the meter on by pressing ON/OFF.
2• When the LCD displays
"- - -", the meter is ready.
3• Fill the cuvet with 10 mL
of unreacted sample, up to
the mark, and replace the
cap.
4• Place the cuvet into the
holder and ensure that the
notch on the cap is
positioned securely into the
groove.
5• Press ZERO and "SIP" will
blink on the display.
6• After a few seconds the display
will show "-0.0-". The meter is
now zeroed and ready for measurement.
7• Remove the cuvet. Open it.
8• Add the content of one packet of
reagent.
Use
MI526-100 for free chlorine or
MI524-100 for total chlorine
9• Replace the cap and swirl
gently for 20 seconds.
Reinsert the cuvet into the
holder and ensure that the
notch on the cap is
positioned securely into
the groove.
10• Wait 1 minute for free chlorine or 2,5
minutes for total chlorine and press
READ. In both cases "SIP" will blink
during measurement.
11• The instrument directly displays
concentration in mg/L of free or
total chlorine.
Interferences:
• Bromine (positive error)
• Chlorine dioxide (positive error)
• Iodine (positive error)
• Oxidized Manganese (positive error)
• Chromium (positive error)
• Ozone (positive error)

Advertisement

loading
Need help?

Need help?

Do you have a question about the MI 404 and is the answer not in the manual?

Questions and answers

Summary of Contents for Milwaukee MI 404

  • Page 1 MI526-100 for free chlorine or necessary information for the correct use of 2• When the LCD displays MI524-100 for total chlorine MI 404 the nstrument. Please read it carefully “- - -”, the meter is ready. before using the meter. This instrument is in 3•...
  • Page 2 • Alkalinity above 250 mg/L CaCO or acidity “-bA-”, the battery is dead and On sample reading BATTERY REPLACEMENT above 150 mg/L CaCO will not reliably must be replaced. Once develop the full amount of color or it may this indication is displayed, •...