Daily; Monthly - Kodak 9000 User Manual

3d extraoral imaging system
Hide thumbs Also See for 9000:
Table of Contents

Advertisement

Chapter 6
MAINTENANCE
This section describes the maintenance tasks that you need to perform regularly for your
KODAK 9000 3D Unit and the accessories.

Daily

Carry out the following maintenance tasks:
Table 6–1
Daily Maintenance Tasks
Accessories
3D and Panoramic bite block
Edentulous bite block
Temple support
Chin rest (panoramic, sinus and
TMJ)
All components that come into
contact with the patient and the
operator
Outer covers of the Unit

Monthly

Wipe the outer covers of the unit with a soft, dry cloth.
WARNING
Switch off the Unit, then, clean all accessible parts of the machine
with an alcohol-based non-corrosive product. Avoid using liquids
inside the Unit. Follow the alcohol-based product manufacturer
recommendations for safety precautions.
CAUTION
You can use the usual disinfectant products, but we recommend
that you protect the Unit from contamination by using barriers
available from dental distributors. Follow the disinfectant
product manufacturer recommendations for safety precautions.
Sterilize with cold sterilization or autoclave up to 134°C before the next
patient is x-rayed.
Sterilize with cold sterilization or autoclave up to 134°C before the next
patient is x-rayed.
Sterilize the head support and chin rest with medical-grade 76% alcohol
disinfectant before the next patient is x-rayed.
Sterilize all components with medical-grade 76% alcohol disinfectant
before the next patient is x-rayed.
Wipe the Unit with a dry cloth at the end of each day's operation.
WARNING
Do not use detergents or solvents to clean the outer covers of the
Unit.
Maintenance Tasks
KODAK 9000 3D Extraoral Imaging System_User Guide (SM710)_Ed 01
6–1

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sm710

Table of Contents