CAPTURE HOOD
MODEL 6710
INSTRUCTION MANUAL
Be sure to read this manual thoroughly before using the instrument.
Fully understand and pay attention to each caution mentioned.
Please keep this manual as a service reference.
01001
1 2 .10
1.800.561.8187
information@itm.com
www.
.com

Advertisement

Table of Contents
loading

Summary of Contents for Kanomax 6710

  • Page 1 CAPTURE HOOD MODEL 6710 INSTRUCTION MANUAL Be sure to read this manual thoroughly before using the instrument. Fully understand and pay attention to each caution mentioned. Please keep this manual as a service reference. 01001 1 2 .10 1.800.561.8187 information@itm.com www.
  • Page 2 Thank you for your purchasing a KANOMAX instrument. This is a precision instrument, please read this instruction manual carefully before using the instrument and pay attention to all the cautions under the instructions. 1.800.561.8187 information@itm.com www. .com...
  • Page 3 Composition & Parts Standard part: Model No. Functions 6710 Air quantity/temperature/humidity Standard components: Items Main structure User Manual Carrying Case Instrument Base 2ft×2ft Fabric hood Frame poles for 2ft×2ft hood Communication cable Software Installer Optional components: Items Specifications Frame poles for other hood sizes...
  • Page 4 ATTENTIONS & NOTES The warning symbols mentioned in this manual are defined below: 〔symbol classifications〕 Danger: To Prevent Serious Injury or Death Warnings in this classification indicate a danger that may result in serious injury or death if not observed. Caution: To Prevent Damage to the Product Warnings in this classification indicate a risk of damage to the product that may void the product warranty if not observed.
  • Page 5 DANGER Do not exposure the instrument to rain or water. …… Otherwise, it may cause an electric shock, fire or injury. Forbidden CAUTION Always unplug the instrument when it is not in use. …… Failure to do so may cause an electric shock, fire or circuit damage. Remove the batteries when storing the instrument for a long period of time.
  • Page 6: Table Of Contents

    CONTENTS 1.Introduction ........................1 1.1 Product features ......................1 2. Capture Hood Structure ....................... 2 2.1 Capture Hood Base ....................2 2.2 Main structure ......................3 2.3 Fabric Hood ........................ 3 3. Installation and Assembling ....................3 3.1 Capture Hood Installation ................... 3 3.2 Assembling the Frame (detail) ..................
  • Page 7 5.1.4 Actual flow, Standard flow settings ..............17 5.2 General settings ....................... 18 5.3 Data processing settings ..................19 5.4 Communication settings ................... 20 5.5 Print Settings ......................21 6. Main Specifications ......................23 7. Common Trouble Shooting ....................24 8.
  • Page 8: 1.Introduction

    1.Introduction The model 6710 is used to measure the air flow at the supply and the exhaust points of an HVAC system. There are several interchangeable hoods of various sizes so it’s easy to find one that will match the size of the supply and exhaust points of any HVAC system.
  • Page 9: Capture Hood Structure

    Capture Hood Structure The capture hood consists of a fabric hood, main structure and an instrument base. The structural drawing is below: Fabric hood Instrument base Main structure 2.1 Instrument Base 1) The exterior structure of the base as below: 2)The internal structure of the base as below: Back pressure plate Portable handle Back pressure on/off...
  • Page 10: Main Structure

    2.2 Main structure Battery Compartment Cover lock Keypad Display LCD Cover Socket POWER   communicatio   Side view Back Front view 2.3 Fabric Hood The fabric hood kit consists of a frame, and a fabric hood with four windows as shown below: Window Frame Fabric hood...
  • Page 11: Assembling The Frame (Detail)

    2)The figures below are provided as a reference. a. Use the mark ① and the stitching as a guide to ensure the hood position. b. Make sure the hood fully wraps around the base. ① 3.2 Assembling the Frame (detail) Insert one end of a support pole into its pole mount in the base as in figure ①...
  • Page 12: Installing The Main Structure

    3.4 Installing the Main structure To install the Main structure ①, set the bottom of the panel in the holder on the unit base ② and gently push the top of the panel backwards ③ until it locks into place. To remove it, simply reverse the procedure.
  • Page 13: Supplying Power With Batteries

    4.2 Supplying Power with Batteries 4 AA-size batteries can be used for a power supply. Press the compartment cover lock (refer to figure ①), and then slide the cover off to open the battery compartment. Put 4 AA-size batteries in and make certain the batteries are correctly oriented. New alkaline battery or rechargeable Ni-MH battery can be used.
  • Page 14: On/Off

    4.3 ON/OFF 1. Getting started The LCD is displayed below for reference: (1)Press【POWER】for a while. (2)Then you enter into the main interface as shown in the left figure. 2. Shut down Press and hold 【POWER】for more than 2 seconds in any mode, and the instrument will turn off automatically.
  • Page 15: Incline Adjustment Settings Of The Instrument

    SAVE: Save the current test results. SET: Store changes or enter into the selected option’s next level interface. MENU: On the main screen holding this down for 2 seconds will enter into the main “MENU”. OK: Pressing for 2 seconds will save the date or time in the “General Settings”, Pressing for 2 seconds will delete saved data in the “Record Processing”.
  • Page 16: Back Pressure Compensation On / Off Settings

    Right switch makes the panel tilt up. Left switch makes the panel tilt down. Note: 1. The instrument may turn off, when you attempt to adjust the incline settings of the instrument if the battery display is at 2. You can’t adjust the incline settings of the instrument when testing or if you’re not in the main interface.
  • Page 17: Measuring

    4.7 Measuring 4.7.1 Start and enter into the main interface To start measuring, go to the main interface menu as shown below: 2012 / 06 / 01 13 : 10 : 10 Cycle : 050 Test ID : 13 Test Mode: Single - - - - SAVE...
  • Page 18: Running Average Measurement Mode

    1)“Single” measurements are individual flow measurements. When you press “START”, the instrument will begin measuring and the air flow value will be displayed as “- - - -”. 2)The test will stop automatically when the air flow is stable and a flow value with the corresponding wind direction will be displayed.
  • Page 19: Back Pressure Compensation Measurement Mode

    4.7.4 Back Pressure Compensation Measurement Mode When measuring air flow in an environment with a strong wind or turbulent flow, it will produce resistance for test air flow and system pressure. It is suggested you turn the Back Pressure Compensation ON when the air flow is more than 1500 m /h, otherwise you can test without the Back Pressure Compensation ON.
  • Page 20: Test Settings

    5.1 Test Settings MENU to select “Test Settings”, and press (1)Press 1.Test Settings MOVE 2. General Settings 3. Record Processing enter into “Test Settings”. Press return to the main 4. Communication 5. Print interface. EXIT (2) The option items in “Test Settings” are as follows: 1.
  • Page 21: Test Id Settings

    5.1.1 Test ID setting A Test ID consists of a group of Cycles (or Samples). A Test ID can contain up to 100 Cycles. The maximum number of Test IDs is 30. Any new Sample will be saved to the current Test ID. You can change the current Test ID at any time to keep your data organized.
  • Page 22: Test Mode Settings

    5.1.2 Test Mode settings This setting is for change the test mode & the settings about the test modes. Refer to the steps below: Test Settings Under “Test Settings”, press to select 2 “Test 1. Test ID MOVE 2. Test Mode Single Mode”.
  • Page 23: Set Units Of Measure For Airflow, Temperature, And Atmos

    5.1.3 Set Units of Measure for Airflow, Temperature and Atmos For available units, refer to the below chart: Display shows Available units Air flow m³/h, CFM, l/s Temperature ℃, ℉ Atmos kPa, inHg To change the units of measure, follow these steps: Test Settings Under “Test Settings”, select the option 3 “Unit”.
  • Page 24: Actual Flow, Standard Flow Settings

    5.1.4 Actual flow, Standard flow settings This setting is for selecting Actual air flow [ACT] or Standard air flow [STD]. Test Settings 1. Test ID Under the “Test Settings”, select option 4 “Act. Std”. MOVE 2. Test Mode Single 3. Unit .
  • Page 25: General Settings

    5.2 General Settings Under “MENU”, select option 2 “General Settings”. MENU 1.Test Settings Press to enter “General Settings” or press MOVE 2. General Settings 3. Record Processing 4. Communication return to the main interface. 5. Print EXIT The settings includes: Date, Time, Backlight and Auto General Settings 1.
  • Page 26: Data Processing Settings

    5.3 Data processing settings MENU 1.Test Setting Under “MENU” select option 3 “Record Processing”. MOVE 2. General Settings 3. Record Processing 4. Communication Press to select “Record Processing”. Press 5. Print return to the main interface. EXIT Record Processing 1.Display MOVE Press to select option 1 or 2.
  • Page 27: Communication Settings

    Under option 1, press Detele Press to select option item as below. 1. Delete All 2. Delete IDs Select ID: Press for 2 seconds to save the changes or press 3. Delete Cycles :12 Select ID to quit. Select Cycle: BACK Under option 2, press Detele...
  • Page 28: Print Settings

    5.5 Print Settings Note:Before printing ensure the following: 1. Please select the factory configuration printer and printer cable. For details, refer to the printer manual <<RD series of mini-printer>>. 2. Baud Rate is set to 19200. The initial value is set to 19200. MENU Under “MENU”, select option 5 “Print”.
  • Page 29 Under option 1, press Print By IDs Press to select the “Start ID” value. Start ID: Press to save or press to quit. Stop ID: Print: Total ID: BACK Print By IDs Press to select the “Stop ID” value. Start ID: Press to save or press to quit.
  • Page 30: Main Specifications

    Main Specifications Items 6710 Flow range 40~4300 m³/h Air Flow Accuracy Readings ±3%± 10 m³/h(Air Flow >85 m³/h) Resolution 1 m³/h Temperature 0~60℃ range Temperature Accuracy ±0.5℃(within 5~40℃±0.3℃) Resolution 0.1℃ Humidity range 0~100%RH Humidity Accuracy ±3%RH(10~90%RH) Resolution 0.1%RH Operating Temperature range 0~60℃(no condensing)...
  • Page 31: Common Trouble Shooting

    7. Common Trouble Shooting Symptom Possible causes / troubleshooting Check the positive/negative connections of the batteries. →Verify the batteries are inserted correctly. No display Low battery charge. →Replace or recharge the batteries. Low battery charge. →Replace or recharge the batteries. A test is running.
  • Page 32: Warranty And After-Service

    To schedule a repair or calibration, please contact the factory. Information for After service * Product Name Capture Hood * Model 6710 * Product Series. ---------- * Symptom detail descriptions * Order date Year Month 1.800.561.8187...
  • Page 33 SHENYANG KANO SCIENTIFIC INSTRUMENT CO., LTD. #2610, 51 Wulihe Street, Heping District, Shenyang PRC. TEL: 024-23846440 83951688 83951788 FAX: 024-23898417-820 URL: http://www.kanomax.com.cn KANOMAX INSTRUMENT(SHENYANG), INC. #9, Zheng kun Road, Shenbei New District, Shenyang, PRC. TEL: 024-89730178 FAX: 024-89730177 □ KANOMAX USA, INC.