berlinger Fridge-tag 2 Operation Manual

With internal sensor
Hide thumbs Also See for Fridge-tag 2:

Advertisement

OPERATION MANUAL
ENGLISH
GEBRAUCHSANWEISUNG
DEUTSCH
Fridge-tag
with internal sensor
PAGE 1-36
SEITE 37-71
®
2

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Fridge-tag 2 and is the answer not in the manual?

Questions and answers

Kelly Cardinal
April 8, 2025

hi, my fridge tag 2l has a triangle with exclamation point, but still has a check mark, How do I correct this?

1 comments:
Mr. Anderson
April 9, 2025

The triangle with an exclamation point on the Fridge-tag® 2 is a warning symbol. If the device still shows a check mark (OK symbol), it means that although a warning condition occurred, the overall status is still acceptable.

To clear the warning symbol:
- Read the details of the triggered alarm or warning from the display.
- After viewing, the warning symbol and any alarm symbol (if present) will disappear, and the display will return to the OK symbol.

Note: The behavior of the symbols is based on the device’s configuration and cannot be reset manually.

This answer is automatically generated

Subscribe to Our Youtube Channel

Summary of Contents for berlinger Fridge-tag 2

  • Page 1 ® Fridge-tag with internal sensor OPERATION MANUAL ENGLISH PAGE 1-36 GEBRAUCHSANWEISUNG DEUTSCH SEITE 37-71...
  • Page 2: Table Of Contents

    Content Page 1) Display explanations 2) State of delivery / Sleep Mode 3) Gathering information prior to device activation 4) Activation process 4.1) Overview sequences of activation 4.2) Activation 4.3) Setting the calendar format 4.4) Use of READ and SET 4.5) Setting the date 4.6) Setting the time 4.7) Setting the alarm limits (not standard)
  • Page 3: Display Explanations

    1) Display explanations Sensor Ext. Int.Sensor 1 - 888 max. 8888 °F min. °C OK ( ) or ALARM (X) indicator HIGH / LOW alarm indicators (showing history of the last 30 days) Power on indicator (double point is flashing) Battery low indicator (approx.
  • Page 4: State Of Delivery / Sleep Mode

    2 is shipped in its so-called „Sleep Mode“. >+8°C (10hrs) <-0.5°C (60min) Temperatur / Min. - Max Alarm state Time / Alarm duration Fridge-tag 2 ® READ =ALARM BAAA00001 The display (LCD) is blank. 3) Gathering information prior to device activation (in Sleep Mode)
  • Page 5 Pressing the READ-button Displayed Information 1st press of READ: Display test: Ext. Int.Sensor Sensor 1 - 888 max. 8888 all segments activated °F min. °C 2nd press of READ: Indication of date and production test result: 16. September 2012 / PASS P SS 3rd press of READ:...
  • Page 6: Activation Process

    4) Activation process 4.1) Overview sequences of activation State of delivery 4.2) Activation 4.3) Setting the calendar format 4.5) Setting the date 4.6) Setting the time 4.7) Setting the alarm limits (only if factory preset otherwise skipped) 4.8) LOC function NOTE: If the activation process has not been completed - after approx.
  • Page 7: Activation

    3 seconds. >+8°C (10hrs) <-0.5°C (60min) Temperatur / Min. - Max Time / Alarm duration Alarm state Fridge-tag 2 ® READ =ALARM BAAA00001 Successful activation is visible when the following indication will appear on the screen: 4.3) Setting the calendar format Option 1: Setting the date format to: dd.mm.yyyy...
  • Page 8: Use Of Read And Set

    Option 2: Setting the date format to: mm.dd.yyyy >+8°C (10hrs) <-0.5C (60min) Temperatur / Min. - Max Alarm state Time / Alarm duration Fridge-tag 2 ® READ Press READ =ALARM BAAA00001 Then press SET to to change save the calendar...
  • Page 9: Setting The Date

    16th of Sep. 2012 (16.09.2012) in Europe format The 1st digit is flashing: >+8°C 10hrs) <-0.5C 60min) Temperatur / Min. - Max Alarm state Time / Alarm duration Fridge-tag 2 ® READ =ALARM Press READ BAAA00001 until „1“ appears as the Press SET first digit...
  • Page 10 The 3rd digit is flashing: >+8°C (10hrs) <-0.5C (60min) Temperatur / Min. - Max Alarm state Time / Alarm duration Fridge-tag 2 ® READ =ALARM Press READ BAAA00001 until „0“ appears as the Press SET third digit to save The 4th digit is flashing: >+8°C...
  • Page 11: Setting The Time

    The 8th digit is flashing: >+8°C (10hrs) <-0.5C (60min) Temperatur / Min. - Max Alarm state Time / Alarm duration Fridge-tag 2 ® READ =ALARM BAAA00001 Press READ until „2“ 2 Press SET appears as the to save eighth digit The date is now set to: 16.09.2012...
  • Page 12 The 2nd digit is >+8°C (10hrs) flashing: <-0.5C (60min) 1 00 Temperatur / Min. - Max Time / Alarm duration Alarm state Fridge-tag 2 ® Press READ READ =ALARM three times BAAA00001 until „3“ 2 Press SET appears as the...
  • Page 13: Setting The Alarm Limits (Not Standard)

    The 1st digit of the duration of -0.5C 0min) the alarm limit is flashing: 0 00 Temperatur / Min. - Max Alarm state Time / Alarm duration Press READ to Fridge-tag 2 ® READ =ALARM adjust the number. BAAA00001 Press SET to confirm the number +8°C 0hrs) -0.5C...
  • Page 14 0.5C min) the alarm limit is flashing: 0 00 Temperatur / Min. - Max larm state Time / Alarm duration Press READ to Fridge-tag 2 ® READ =ALARM adjust the number. BAAA00001 2 Press SET to confirm the number >+8°C...
  • Page 15 0 °C / 0 °F and +50 °C / +122 °F >+8°C (10hrs) <-0.5C (60min) °C Temperatur / Min. - Max Alarm state Time / Alarm duration Fridge-tag 2 ® READ Press READ until the =ALARM BAAA00001 display shows Press SET to adjust the no flashing sign: limit between 0 °C / 0 °F...
  • Page 16 >+8°C (10hrs) <-0.5C (60min) - 80 °C Temperatur / Min. - Max Time / Alarm duration Alarm state Fridge-tag 2 ® READ =ALARM Press READ until the BAAA00001 flashing „ “ sign is Press SET to set the indicated on the screen: limit below 0 °C / 0°F...
  • Page 17: Loc Function

    (10hrs) <-0.5C (60min) 1 47 Temperatur / Min. - Max Time / Alarm duration Alarm state Fridge-tag 2 ® READ NOTE: =ALARM • Even in LOC mode the user can press the READ button to BAAA00001 retrieve history information, change any setting or download a report to a computer.
  • Page 18: Placing The Fridge-Tag 2

    Note: For the right positioning of the Fridge-tag 2 within the ® fridge, please follow the instructions of WHO, CDC or any other governmental requirements of your country. (For more information visit www.berlinger.com)
  • Page 19: Read And Change Settings / How To Correct Setting Mistakes

    Read and change settings / How to correct setting mistakes 6.1) Overview menu Device activated 6.2) Menu entry 6.2.1) Menu: SET DATE (Change date and / or time) 6.2.2) Menu: READ CONF (Read alarm configurations) 6.2.3) Menu: CELS FAHR (change alarm configurations) 6.2.4) Menu: SET CONF (only if factory preset otherwise skipped) 4.8) LOC function...
  • Page 20: Menu Entry

    To change the date format, the date, the time, the temperature measurement unit or the alarm settings or to read the pre-set alarm limits please proceed as follows: Temperatur / Min. - Max arm state Time / Alarm duration Fridge-tag 2 ® READ =ALARM BAAA00001 Press and continue to ...then press...
  • Page 21 Information: • Time and date adjustments have no effect on the alarm records. • Once the device is activated, it cannot be stopped anymore. • The number of time adjustments during the same day is unlimited. Note: After the adjustment has been done, the Fridge-tag 2 will be locked for 24 hours from the following ®...
  • Page 22: Display Indication During Measurement

    Int.Sensor max. 0 00 °C Temperatur / Min. - Max Time / Alarm duration Alarm state Fridge-tag 2 ® READ =ALARM BAAA00001 During max. 1 minute after the LOC-period no current temperature is displayed on the screen. Example of OK Display - during measurement >+8°C...
  • Page 23: Warning Symbol

    Int.Sensor 1 42 °C Temperatur / Min. - Max Alarm state Time / Alarm duration Fridge-tag 2 ® READ If the temperature and time conditions are outside the preset alarm =ALARM parameters the following will be displayed on the screen: BAAA00001 •...
  • Page 24: Option 1: Read Out Day-Per-Day

    Int.Sensor max. 0 32 °C Temperatur / Min. - Max Alarm state Time / Alarm duration Press READ once Fridge-tag 2 ® READ =ALARM The following information is BAAA00001 indicated on the screen: • The OK symbol • The corresponding arrow (example: high arrow of „today“) •...
  • Page 25 >+8°C (10hrs) <-0.5C (60min) 1 21 Temperatur / Min. - Max Alarm state Time / Alarm duration Fridge-tag 2 ® READ =ALARM BAAA00001 Press READ The following information is indicated on the screen: • The ALARM symbol • The corresponding arrow (example: Lower ALARM limit) •...
  • Page 26: Option 2: Read Out Only Alarms (Use Alarm-Super-Jump Function)

    >+8°C (10hrs) <-0.5C (60min) 2 30 Temperatur / Min. - Max Time / Alarm duration Alarm state Fridge-tag 2 ® READ =ALARM BAAA00001 Press READ for 3 seconds The following information is indicated on the screen: • The ALARM symbol •...
  • Page 27 (60min) Int.Sensor max. 1 24 Temperatur / Min. - Max Alarm state Time / Alarm duration Fridge-tag 2 ® READ =ALARM BAAA00001 Then press READ again The following additional information is indicated on the screen: • Highest recorded temperature (example: +10.5 ˚C) •...
  • Page 28 9.3) Option 3: Read out data from the files generated by the Fridge-tag ® by connecting it with a computer (60 day history) Plug the Fridge-tag 2 into any computer via USB Interface. Make ® sure the device is plugged in properly. The following window will appear: Wait sufficient time for the device to generate the ASCII and PDF files (approx.
  • Page 29 Note: For a proper USB-port disconnection of the device, please always use the “safely remove hardware” function on your PC/Mac. Right mouse click on the icon Eject … (choose the right device to remove). Do not disconnect the device before you receive the following message:...
  • Page 30 Sample of a PDF-file generated by a Fridge-tag ® Page 1 of 2...
  • Page 31 Information: File names on the Fridge-tag 2 are write protec- ® ted. The names may only be changed after downloading onto a computer. Changing is either possible directly on unopened files or via open and save commands with the Adobe Reader. Using other programs may cause loss of the digital signature.
  • Page 32: Veryfication Process

    ® or accidentally changed (meets the strict FDA 21 CFR Part 11 requirements). 1st step: Download the software „Verifier“ from our website: www.berlinger.com 2nd step: Open the software. The following window will appear: 3rd step: Click on „Open file“ 4th step: Select the file you would like to verify.
  • Page 33: Explanation Of Terms

    ® July 2016 (2016-07). More information about the Fridge-tag 2 can be found in the sales ® brochure and on the website: www.berlinger.com 12) Technical specifications Storage condition (inactive) 0 °C to +30 °C Operating temperature -25 °C to +55 °C* Display visible -10 °C to +55 °C...
  • Page 34: Important Information

    13) Important Information Liability The manufacturer shall not be held liable: • if the device was used beyond the manufacturer’s given limitations. • for any claims due to the improper storage and use of the device. • for any problems with the temperature controlling and / or cooling unit. •...
  • Page 35 Manufacturer: Q-tag AG Mitteldorfstrasse 2 9608 Ganterschwil Switzerland www.q-tag.com Sales: Berlinger & Co. AG Mitteldorfstrasse 2 9608 Ganterschwil Switzerland Tel.: +41 71 982 88 11 Fax: +41 71 982 88 39 E-Mail: info@berlinger.com Web: www.berlinger.com 20151126_Release_10.1 Subjet to change. Please note that all information in this document is correct at the time of publication. Due to our policy of continuous product development, we reserve the right to change this information without prior notice.

Table of Contents