(e)
Ensure that all the cleaning fluid has evaporated before replacing the metal
sensor sealing cap assembly. Take care to ensure that the "o" ring seal is
correctly seated.
(f)
After assembly, press the INITIATE push-button to initiate an ABC cycle.
Then re-balance the instrument.
In a clean environment it is quite normal for the instrument to operate satisfactorily for
several months without the need for optics cleaning. However, it is good practice to
ensure the cleaning procedure is carried out regularly rather than waiting for the red ABC
STATUS LED to illuminate.