Appendix
A.
Specification List
DC Characteristics
Function
Range
100.0000
mV
1.000000 V
DCV
(DC
10.00000 V
Voltage)
100.0000 V
1000.000 V
Function
Range
10.00000
mA
100.0000
DCI
mA
(DC
Current)
1.000000 A
3.00000 A
24 hours accuracy
Input
(% of reading+% of
Resistance
range)(23
Ω
>10 G
0.0030+0.0030 0.0050+0.0035
Ω
>10 G
0.0020+0.0006 0.0040+0.0007
Ω
>10 G
0.0015+0.0004 0.0035+0.0005
Ω
10 M
0.0020+0.0006 0.0045+0.0006
Ω
10 M
0.0020+0.0006 0.0045+0.0010
24 hours accuracy
Shunt
(% of reading+% of
Resistance
range)(23
Ω
10.1
0.005+0.010
Ω
10.1
0.01+0.004
Ω
0.1
0.05+0.006
Ω
0.1
0.10+0.020
89
±
1 year accuracy
(% of reading+% of
±
o
o
o
C
1
C)
range)(23
C
±
1 year accuracy
(% of reading+% of
±
o
o
o
C
1
C)
range)(23
C
0.050+0.020
0.050+0.005
0.100+0.010
0.120+0.020
±
±
o
5
C)
±
±
o
5
C)