Transformer - Picotest M3500A Service Manual

6.5 digit digital multimeter
Hide thumbs Also See for M3500A:
Table of Contents

Advertisement

2.1.2 Transformer

Usually you can change the line voltage setting by voltage selector as the
shown in section 2.1.1. If the voltages shown in the selector do not fit your
requirement, you may change the transformer connections to change the
line voltage setting selections. The voltages can be set for M3500A are 100,
120, 220, 240 Vac.
Warning! Do not perform this action unless you are qualified to do so.
For more information, please contact your local service representative.
[Step 1]
Turn off the power and disconnect the line cord from your multimeter as
shown in Figure 2-5. Also, you should remove all the leads connected to it.
[Step 2]
Remove the handle and then the metal cover. Please refer to section 4.2 for
more detail.
Warning! Do not open the case unless disconnecting the AC line cord
and all test leads. Do not perform this action unless you are qualified to do
so.
Figure 2-5
11

Hide quick links:

Advertisement

Table of Contents
loading

Related Products for Picotest M3500A