1. Warnings

Your new Ultrasonic Liquid Processor has been designed, built and tested to assure
maximum operator safety. However, no design can completely protect against improper use
that may lead to bodily injury and/or property damage. For total safety and equipment
protection, read the instruction manual carefully before attempting to operate this equipment.
Observe the following WARNINGS:
High voltage is present in the generator (power supply), converter and high frequency cable. There are no
user-serviceable parts inside any of these devices. Do NOT attempt to remove the generator cover or
converter case.
Do NOT touch any open cable connections on the unit while the power is turned ON.
Do NOT operate generator with converter disconnected from high voltage cable. High voltage is present in
the cable and may pose a shock hazard.
Do NOT attempt to disconnect the converter high voltage cable while the unit is running.
The generator must be properly grounded with a 3-prong plug. Test electrical outlet for proper grounding
before plugging in unit.
Install the SONICATOR
protected from extremes in temperature and humidity. Do not place the Generator within a Fume Hood.
NEVER immerse the converter in liquids of any kind, or let condensed moisture or liquid drip into the
converter.
NEVER grasp an activated horn or touch the tip of a vibrating probe. It can cause severe burns and tissue
damage.
NEVER hold or clamp the converter by the front driver (black portion) or by the horn itself. This can cause
permanent damage to the system. Support the converter by only clamping around the converter case.
Do NOT allow the tip of a vibrating horn or probe to touch the counter top or any other hard surface. It could
damage the probe, overload the generator, or damage the surface.
Avoid touching the bottom or sides of a glass or plastic container with an activated probe. It could crack or
shatter the glass or melt the plastic.
Turn OFF the power switch, unplug the generator and disconnect the power cord from the back of the
generator before attempting to replace the fuses.
Inspect high frequency cable for cracks in the protective outer jacket.
Do not operate unit with a damaged cable. Doing so may cause serious injury.
In case of AC power loss, wait 3 minutes minimum before reapplying power.
Do not turn off AC mains power while running a horn. Stop sonication through the touch screen prior to
removing AC power.
Symbols
Caution, Risk of electric shock, Hazardous voltage.
®
in an area free from excessive dust, dirt, explosive or corrosive fumes and