Specification data

All accuracy statements are for one year.
Specification - General
Languages
English [Default]
Operating
14 ... 122°F (-10 ... 50°C)
temperature
Storage
-4 ... 158°F (-20 ... 70°C)
temperature
Humidity
0 to 90% without condensation
(Def Stan 66-31, 8.6 cat III)
Shock/Vibration
BS EN 61010:2001; Def Stan 66-31, 8.4 cat III
Specification - Pressure measurement
The % full scale (FS) statements for Standard accuracy and
Premier precision are only applicable if there is a regular
zero correction by the IDOS instrument.
Ranges: Gage and differential operation (g/d),
sealed gage (sg), absolute (a)
+/- psi: 0.36 (+/- mbar: 25)
+/- psi: 1, 3, 5, 10 (+/- mbar: 70, 200, 350, 700)
psi: -15 to [15 or 30]
(bar: -1 to [1 or 2])
psi: -15 to [50, 100, 150, or 300]
(bar: -1 to [3.5, 7, 10, or 20])
psi: 500, 1000, 1500, 2000, 3000
(bar: 35, 70, 100, 135, 200)
psi: 5 (mbar: 350)
psi: 30 (bar: 2)
psi: 100, 300 (bar: 7, 20)
psi: 5000, 10000 (bar: 350, 700)
*
Standard accuracy is for 32 ... 122°F (0 ... 50°C);
Stability: 1 year
Premier precision is for 65 ... 82°F (18 ... 28°C);
Stability:
10 psi (700 mbar) = 0.02% of reading/Year
>
Stability:
10 psi (700 mbar) = 0.01% of reading/Year
Premier precision for 41 ... 113°F (5 ... 45°C):
10 psi (700 mbar): 0.075% FS
>
10 psi (700 mbar): 0.014% FS
K335 Issue 3
EMC
Safety
Size (L: W: H)
Weight
Power supply
Duration
(Measure)
Type
Standard Accuracy *
% FS
g/d
0.1
g/d
0.075
g/d
0.05
g/d
0.05
g/d
0.05
a
0.1
a
0.075
a
0.075
sg
0.05
Notes:
1.
Reference port media: Non-corrosive, dry gas
2.
+ port media: Non-corrosive, non-conductive liquid or
Non-corrosive, dry gas
3.
+ port media: Media applicable to stainless steel
BS EN 61326-1:1998 + A2:2001
Electrical - BS EN 61010:2001;
Pressure Equipment Directive - Class: Sound
Engineering Practice (SEP); CE Marked
Maximum: 7.5 x 3.3 x 2.0 in
(190 x 85 x 50 mm)
19 oz (530 g)
3 x AA alkaline batteries
P1: ≈ 50 hours
P1 + mA: ≈ 5 hours (24 V Source at 12 mA)
Premier Precision †
% FS
0.03
0.03
0.01
0.01
0.01
-
-
-
-
[EN] English -
Notes
1/2
1/2
1/2
1/3
1/3
2
2
3
3
11