4.3 Removing a piston
IMPORTANT (NOTE)
Perform this procedure if the piston is seized.
DANGER – Serious damage to health / risk to life
Isolate the analyzer from mains power supplies before
performing this procedure.
WARNING – Bodily injury
— do not allow any fluid from the measurement cell
sample tube to come into contact with either the
skin or any metallic / electronic parts of the
analyzer. The fluid is contaminated with acid from
the reagents used in the analyzer – wipe up any
spillages immediately
– wear appropriate PPE when emptying any liquid
from the piston
Referring to Fig. 4.2:
1. Turn handle A
1
/
turn clockwise, pull the analytical
4
section forward and allow it to rest against its stops.
Fig. 4.2 Opening the analytical section
6
SI/ANAINST/001-EN | Aztec 600 | Piston assembly | Single- and multi-stream colorimetric analyzers
Referring to Fig. 4.3:
2. Disconnect the measurement cell ribbon cable B from
the connector on the back of the measurement cell.
3. Remove 4xM4 screws C and remove the measurement
cell assembly.
Fig. 4.3 Removing the measurement cell assembly
Referring to Fig. 4.4:
4. Depress catch D and remove cover E.
Fig. 4.4 Removing the measurement cell cover
Referring to Fig. 4.5,
5. Invert the measurement cell assembly, remove 4 screws
F and remove base plate G.
6. Remove sample tube end face sealing cap H and remove
seal I. Discard the seal.
7. Remove all traces of fluid from the inside of sample tube
J.
Fig. 4.5 Removing the measurement cell base plate