Error Information
Detector Error Messages

Grating or Filter Motor Errors

Error ID: Grating: 7800, 7801, 7802, 7803, 7804, 7805, 7806, 7808, 7809;
Filter: 7810, 7811, 7812, 7813, 7814, 7815, 7816
The motor test has failed.
Test 0 Failed:
Test 1 Failed:
During the motor tests, the detector moves the motor to the end position
while monitoring the end- position sensor. If the end position is not found,
the error message is generated.
Probable cause
1
2
3
4
90
Motor is not connected.
Defective motor.
Defective/missing grating or filter.
Cable/connector defective.
Filter motor.
Grating motor.
Suggested actions
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.