Agilent Technologies 1290 Infinity II User Manual page 175

Variable wavelength detector
Hide thumbs Also See for 1290 Infinity II:
Table of Contents

Advertisement

Table 14
Pin
1
2
3
4
5
6
7
8
Special Interfaces
There is no special interface for this module.
Agilent 1290 Infinity II VWD User Manual
ERI signal distribution
Signal
Description
START REQUEST
(L) Request to start injection cycle (for example, by start key on
any module). Receiver is the autosampler.
STOP
(L) Request to reach system ready state as soon as possible (for
example, stop run, abort or finish and stop injection). Receiver is
any module performing run-time controlled activities.
READY
(H) System is ready for next analysis. Receiver is any sequence
controller.
POWER ON
(H) All modules connected to system are switched on. Receiver is
any module relying on operation of others.
Not used
SHUT DOWN
(L) System has serious problem (for example, leak: stops pump).
Receiver is any module capable to reduce safety risk.
START
(L) Request to start run / timetable. Receiver is any module
performing run-time controlled activities.
PREPARE
(L) Request to prepare for analysis (for example, calibration,
detector lamp on). Receiver is any module performing pre-analysis
activities.
Hardware Information
Interfaces
12
175