Table 3: Guidance and manufacturer's declaration – electromagnetic immunity
The 658 is intended for use in the electromagnetic environment specified below.
The customer or the user of the 658 should assure that it is used in such an environment.
Immunity test
3V,
0.15MHz-80MHz
6V in ISM and
Conducted RF IEC
amateur radio bands
61000-4-6
between 0.15MHz
and 80MHz
10V/m
Radiated RF
80MHz-2.7GHz
IEC 61000-4-3
80% AM at 1kHz
WARNING: Portable RF communications equipment (including peripherals such as antenna cables and external
antennas) should be used no closer than 30 cm (12 inches) to any part of the 658, including cables specified by
the manufacturer. Otherwise, degradation of the performance of this equipment could result.
Proximity fields from RF wireless communications equipment
Test frequency
(MHz)
385
450
710
745
780
810
870
930
1720
1845
1970
2450
5240
5500
5785
IEC 60601
Compliance
Test level
3V,
0.15MHz-80MHz
6V in ISM and
amateur radio bands
between 0.15MHz
and 80MHz
10V/m
80MHz-2.7GHz
80% AM at 1kHz
Band
(MHz)
380-390
430-470
704-787
800-960
1700-1990
2400-2570
5100-5800
Electromagnetic environment
level
Portable and mobile RF communications equip-
ment should be used no closer to any part of the
SYSTEM, including cables, than the recom-
mended separation distance calculated from the
equation applicable to the frequency of the trans-
mitter.
Recommended separation distance
Where P is the maximum output power rating of the
transmitter in watts(W) according to the transmitter
manufacturer and d is the recommended separa-
tion distance in metres (m).
Field strengths from fixed RF transmitters, as de-
termined by an electromagnetic site survey, should
be less than the compliance level in each frequency
range.
Interference may occur in the vicinity of equipment
marked with the following symbol:
Immunity test levels
Professional healthcare facility environment
Pulse modulation 18Hz, 27V/m
FM, ±5kHz deviation, 1kHz sine, 28V/m
Pulse modulation 217Hz, 9V/m
Pulse modulation 18Hz, 28V/m
Pulse modulation 217Hz, 28V/m
Pulse modulation 217Hz, 28V/m
Pulse modulation 217Hz, 9V/m
guidance
80MHz to 800MHz
800MHz to 2.5GHz
Need help?
Do you have a question about the ADC Adscope 658 and is the answer not in the manual?
Questions and answers
Not keeping a charge, I have received a new charging cord about a year ago, starting to not keep a charge. please help
The ADC Adscope 658 may not be keeping a charge due to a poor battery connection or a faulty battery. If the battery is properly connected and still does not charge, it likely needs to be replaced.
This answer is automatically generated