Specifications; General Specifications - YOKOGAWA CW500 Getting Started Manual

Power quality analyzer
Hide thumbs Also See for CW500:
Table of Contents

Advertisement

Specifications

General Specifications

Bluetooth function
Measurement line
Measured and computed items Voltage, current, frequency, active power, reactive
Other functions
Voltage
Range
(RMS)
Accuracy
Effective input range 1% to 120% (rms) of each range and 200% (peak) of
Display range
Crest factor
Voltage transient
sampling speed
Current
Range
(RMS)
Accuracy
Effective input range 1% to 110% (rms) of each range and 200% (peak) of
Display range
Crest factor
IM CW500-02EN
CW500-B0: Not available CW500-B1: Available
Single-phase two-wire (up to four systems), single-
phase three-wire, three-phase three-wire, three-
phase four-wire
power, apparent power, active energy, reactive
energy, volt-ampere hours, power factor, neutral line
current, demand, harmonics, power supply quality
(swell, dip, interruption, transient overvoltage, inrush
current, unbalance factor, IEC flicker)
Digital output, analog DCV input
600.0/1000 V
Measurement waveform
Sine wave 40 to 70 Hz at 600 V range
10% to 150% for nominal voltage 100 V or higher:
nominal voltage±0.5%
Outside the above range and 1000 V range:
±0.2% rdg ± 0.2%rng
each range
0.15% to 130% of each range (displays 0 for less
than 0.15%)
3 or less
24 µs
96060(2A type): 2A
96061 (50A type): 5/50A/AUTO
96062 (100A type): 10/100A/AUTO
96063 (200A type): 20/200A/AUTO
96064 (500A type): 50/500A/AUTO
96065 (1000A type): 100/1000A/AUTO
96066 (3000A type): 300/1000/3000A
±0.2% rdg ± 0.2% rng + current clamp-on probe
accuracy (sine wave, 40 to 70 Hz)
each range
0.15% to 130% of each range (displays 0 for less
than 0.15%)
3 or less
37

Advertisement

Table of Contents
loading

Table of Contents