Power Supply And Output Connections To Remote Transmitter - ABB Endura AZ20 series Maintenance Manual

Probe combustion oxygen monitor
Hide thumbs Also See for Endura AZ20 series:
Table of Contents

Advertisement

Combustion Oxygen Monitor
Endura AZ20 Series Probe

6.3.4 Power Supply and Output Connections to Remote Transmitter

Warning.
The transmitter must be earthed.
Isolate the incoming mains power supply cable before
making connections at the transmitter or the probe.
C
Brown
Blue
L/L1
N/L2
*Refer to the transmitter User Guide (IM/AZ20E–EN)
for HART communication details
Fig. 6.8 Remote Transmitter – Power Supply and Output Connections
IM/AZ20M–EN Rev. B
Earth (Green /
Yellow)
C
C
N/C N/C
G
Mains
+
Connections
C
B
A
D
To make power supply and output connections:
1. Feed the incoming AC power supply cable through cable
A
gland
.
2. At terminal block
supply line (brown) and neutral (blue) terminals
3. Connect the incoming AC power supply earth wire to
internal earth connection
4. Feed the signal cable(s) through cable gland(s)
F
and make connections to the relay outputs (terminal
B
block
) and current output and option board terminals
(terminal block
5. Refer to Section 6.1.1, page 36, steps 5 to 8 to refit the
cartridge and transmitter cover.
Fuse F1 (AC) 1.0 A Type F
100 to 240 V AC (±10 %)
50 / 60 Hz
B
Current Output
(4 to 20 mA) HART*
Option Board
Connections**
**Option Board Connections
Analog output
Digital I/O
6 Dismantling and Reassembly – Transmitter
B
, make connections to the AC power
D
.
G
) as required.
D
A
E
F
A
B
C
+
DIO1
DIO2
COM
C
.
E
and
41