5905/5010 BOD P
SING
HE
Note: The 5905 must be plugged into a YSI meter and the
stirrer power plug must be plugged into its power adapter. The
5010 must be plugged into a YSI meter only, as it gets its
stirring power from the 5000/5100 instrument.
C
M
HANGING
New Probe Use
New probes are shipped with a dry membrane cap to protect
the electrodes. A new membrane cap must be installed
before first use.
Follow these instructions to change membrane caps:
1. Remove the stir paddle from the probe by pulling it straight
out.
2. Unscrew the old membrane cap from the probe. Before
installing a new membrane, clean the probe tip with
deionized water in order to remove any contaminants.
3. Hold the membrane cap and fill it at least ½ full with the
electrolyte solution provided. Screw the membrane cap onto
the probe moderately tight. A small amount of electrolyte
should overflow.
4. Rinse off excess electrolyte from the probe with deionized
water.
5. Reinstall the stir paddle.
C
EMBRANE
APS

ROBE

- F
2
IGURE
3