YSI 5905 Instruction Manual page 14

Bod probe
Table of Contents

Advertisement

Symptoms
Unstable
readings
Motor not
working
Loud motor
Membrane cap is
tight
Actions
Allow sufficient time for
temperature and oxygen
stabilization
Allow 60 sec for reading to
stabilize
Change membrane
Clean probe (see
instructions under
Electrode Cleaning)
Check power supply
Manually turn the stir
paddle to help start the
motor
Replace motor
Check if stir paddle is
pushed all the way in
Replace motor
Use pliers to loosen the cap
(turn counter clockwise
while facing sensor tip)
10