Technical Data; Warranty - TESTO 0636.9835 Application Information

Pressure dew point probe
Table of Contents

Advertisement

12
Sensor:............................................testo
Measuring range: ...........................0 to 100 %RH,
Pressure dew point t
Pressure range: ..............................Overpressure to 15 bar
Accuracy, temperature:* ................±0.4 °C
Resolution: .....................................0.1 %RH / 0.1 °C
Temperature range: .......................Probe: -20 to +50 °C
Control point: .................................11.3 to 75.3%RH
Measuring air flow rate
at 6.0 bar overpressure: ................1,0 l/min factory set at
........................................................delivery (relating to 0 bar,
........................................................+20°C)
Response time: ..............................1 to 5 min
Storage/transport temperature: ....-20 to +70°C
Accuracy:
dew point
* Accuracy referred to a rated temperature of +25 °C.
The Technical data applies in connection with the testo x35
measuring instrument.
Pressure dew point probe ................1 year
Measuring chamber .........................1 year
If the instrument is opened, improperly handled or if force is applied, the
: .....................-60 to +50 °C t
pd
Instrument: 0 to +40 °C
typically 2 min
Pressure
in °C
0636.9835
5
±0.9
0
±1
-5
±2
-10
±3
-20
±4
-30
-
-40
-
warranty will be no longer valid.

Technical data

®
ceramic sensor
pd
Accuracy
in °C t
pd
0636.9836
±0.8
±0.8
±0.9
±1
±2
±3
±4

Warranty

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

0636.9836

Table of Contents