Ne remplacez pas les composants car cela pourrait interférer avec la sécurité intrinsèque de l'appareil. DO NOT substitute any other battery type than specified and supplied by Gas Clip Technologies. Only use Gas Clip Technologies parts in the detector. Non approved parts will void the warranty and are considered unsafe.
Multi Gas Clip User’s Manual Basic Operation Button Usage Operation of the detector is driven by a single button located on the front of the instrument: Turn On/Off Menu Navigation (Status and Options) Latched Alarm acknowledgement ...
Multi Gas Clip User’s Manual Display Components Display Layout Entry Description Alarm Condition Calibration Bottle Battery Level Gas Readings Gas Identifiers Display Details When a gas is reading at, or above its alarm thresholds, the gas identifier icon will flash and the associated alarm condition icon will display.
Multi Gas Clip User’s Manual Alarms Default Alarms Each detector comes preprogrammed with the factory default Low, High, Time Weighted Average (TWA) and Short Term Exposure Limit (STEL) alarms: Sensor High STEL 10 ppm 15 ppm 10 ppm 15 ppm...
Multi Gas Clip User’s Manual Detector Maintenance: Bump Testing The detector can be configured to keep track of regular bump testing intervals. When the detector has exceeded the bump interval, it will display “BUMP DUE” until it has been successfully bump tested. To bump test the unit, either insert the detector into the Clip Dock test station, or manually apply gas from the bump test screen described below.
DO NOT charge the instrument in a combustible atmosphere. DO NOT charge the instrument in temperatures above or below the specified range of 0°C to 40°C DO NOT substitute any other battery type than specified and supplied by Gas Clip Technologies. V1.7 Portable gas detectors you can count on.
Multi Gas Clip User’s Manual Instrument Options/Menus To access the user menus press the button two separate times in quick succession (“double tap”). The detector will display: 1. Current date/time 2. User-programmed text message 3. Status menu prompt 4. Options menu prompt...
Multi Gas Clip User’s Manual Clear Peak Levels: Pressing the button during the “CLEAR ALL” prompt will cause the TWA, STEL and peak readings to be cleared. Note: this does not remove the information from the internal memory logs of the detector.
Page 11
Multi Gas Clip User’s Manual Calibration Interval: The calibration interval controls how often the detector notifies the user to calibrate the sensors. The interval can be individually adjusted for each sensor from 1 to 180 days. Default is 180 days.
Multi Gas Clip User’s Manual Instrument Records (Logs): During operation the instrument keeps an active log of all instrument activity. These records can be downloaded from the instrument via the MGC IR Link software and an IR Link, or from the Clip Dock.
Multi Gas Clip User’s Manual Failures/FAQ’s Err: If a sensor is displaying “Err” this sensor has failed and is therefore disabled. Contact GCT for either warranty replacement or replacement sensor(s). BUMP DUE: If the detector is displaying “BUMP DUE”, the detector is either due for a bump test because of a scheduled test or has failed its last bump test.
Multi Gas Clip User’s Manual Instrument Specifications Size 4.7 x 2.4 x 1.2 in. (118.6 x 61.6 x 31.7 mm.) Weight 7.7 oz. (240 g) Temperature -4 to +122˚F (-20 to +50˚C) Humidity 5% to 95% RH (non-condensing) 60 days continuous...
Multi Gas Clip User’s Manual Contact Information Gas Clip Technologies, Inc. 218 W Avenue F Midlothian, TX 76065 Toll Free: (877) 525-0808 Phone: (972) 775-7577 Fax: (972) 775-2483 E-mail: info@gascliptech.com Website: www.gascliptech.com V1.7 Portable gas detectors you can count on.
Multi Gas Clip products. This warranty extends only to the sale of new and unused products to the original buyer. GCT’s warranty obligation is limited, at GCT’s option, to refund of the purchase price, repair, or replacement of a defective product that is returned to a GCT authorized service center within the warranty period.