Summary of Contents for Gas Clip Technologies Single Gas Clip

  • Page 1 Single Gas Clip Single Gas Clip Plus User’s Manual...
  • Page 2: Table Of Contents

    SGC and SGC Plus User’s Manual CONTENTS Warning Statements/Avertissement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 READ FIRST BEFORE OPERATION .
  • Page 3: Warning Statements/Avertissement

    The CO and H S versions of both the Single Gas Clip and SGC Plus do not require calibration for the life of the product. The detector should be bump tested before each day’s use with a known concentration of gas to confirm its ability to respond to gas.
  • Page 4 Étalonnez le détecteur SGC O₂ au moins tous les 30 jours. Assurez-vous de calibrer dans un environnement d’air propre. Voir la section sur le détecteur SGC O₂ pour des instructions sur l’étalonnage. Les versions CO et H2S du Single Gas Clip et du SGC Plus ne nécessitent pas d’étalonnage pendant toute la durée de vie du produit.
  • Page 5: Detector Components

    SGC and SGC Plus User’s Manual DETECTOR COMPONENTS ENTRY DESCRIPTION Audible Alarm Port Display Alarm Bar LEDs Maintenance LEDs Power Button Sensor Infrared (IR) Communication Window Alligator Clip with Safety Ring Certification Label UM SGC v3.0 3 of 15...
  • Page 6: Display Components

    SGC and SGC Plus User’s Manual DISPLAY COMPONENTS Display Layout ENTRY DESCRIPTION Alarm Condition Self-Test Icon Test Reminder Icon Gas Type Identifiers Battery Indicator (Used During Real-Time Gas reading) Life Remaining or Real Time Gas Reading Data High And Low Alarm Set Point Icon 8/11 Detector Life Remaining Icon Infrared Data Transfer Icon...
  • Page 7: Basic Operation

    After the full element LCD screen, the low alarm and high alarm set points will be displayed. Note: these alarm set points can be adjusted using the GCT IR Link or Single Gas Clip Dock configuration options. Factory Standard Alarm Set Points:...
  • Page 8 SGC and SGC Plus User’s Manual Screen 4: (If applicable): If programmed with a “User ID”, after the alarm set points are displayed, a combination of numbers and or letters will scroll across the screen. This will be a max of 2 screens with a maximum character limit on the “User ID”...
  • Page 9: Alarms

    SGC and SGC Plus User’s Manual ALARMS When a particular gas has a Gas Reading (4) at, or above, its alarm thresholds, the Gas Identifier (5) icon will flash and the associated Alarm Condition (1) icon will display. During a calibration or bump test, the Calibration/Test Mode (2) icon will display when it is time to apply gas. The Battery Charge Level (3) is displayed in a 3 bar battery icon as well as a percentage.
  • Page 10: Detector Maintenance

    GCT recommends users of the O₂ Single Gas Clip to bump test the detector before using each day. Single Gas Clip Oxygen (O₂) detector factory default will prompt the user to calibrate the detector every 30 days. The user will be prompted by the screen flashing CAL, please see calibration instruction below: Calibration Instructions Only perform O₂...
  • Page 11: Cleaning

    SGC and SGC Plus User’s Manual Cleaning The detector can be cleaned with a soft damp cloth. Do not use solvents, soaps or polishes. A neutral cleaner, like Mat & Table Top Cleaner (by ACL Staticide) may also be used. The sensor screen may be cleaned with a soft brush under clean, warm water.
  • Page 12: Detector Records (Logs)

    SGC and SGC Plus User’s Manual DETECTOR RECORDS (LOGS) Event Log By default the detector stores the last twenty-five (25) alarm events. The system stores events by first in first out, i.e. the 26th event will replace the first event and so on. This information can be downloaded using the GCT IR Link. For each alarm event the detector records the following: •...
  • Page 13: Accessories And Replacement Parts

    High Pressure and Ethernet version. • Single Gas Clip Dock Wall Mount: (P/N: SGC-WMDOCK) - Same capabilities as an SGC Dock in a metal case that can be wall mounted or stand alone. - also available in a High Pressure and Ethernet version.
  • Page 14: Detector Specifications

    SGC and SGC Plus User’s Manual DETECTOR SPECIFICATIONS Size 3.3 x 2 x1.1 in (85x50x28 mm) Weight 2.7 oz (76 g) Temperature -40 to 122 F (-40 to +50 C) for CO and H S, -31 to +122 F (-35 to +50 C) for O₂...
  • Page 15 SGC and SGC Plus User’s Manual Conditions of Acceptability: Under certain extreme circumstances, exposed plastic and unearthed metal parts of the enclosure may store an ignition-capable level of electrostatic charge. Therefore, the user/installer shall implement precautions to prevent the buildup of electrostatic charge, e.g.
  • Page 16: Warranty

    (2) years of operational use by the original end user, whichever occurs first. The GCT Product Warranty period for Single Gas Clip Plus monitors shall begin upon the date of activation by the original end user or upon the “Activate Before Date”...
  • Page 17 DAMAGES, ARISING OUT OF BREACH OF THIS WARRANTY OR ANY IMPLIED WARRANTY (INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY). Gas Clip Technologies, Inc. 305 W. FM 1382, Ste. 540, Cedar Hill, TX 75104 Phone: +1 (972) 775-7577 • Toll Free: 1 (877) 525-0808 • Fax: +1 (972) 775-2483 info@gascliptech.com •...