Carbolite 3508 P1 Operating Instructions Manual page 15

Temperature controller eurotherm 3508 series; eurotherm 3504 series;
Table of Contents

Advertisement

Finish the program with an End segment:
Press scroll  until Segment Type is displayed. Press up  or down to select 'End'
11.
Press scroll  until End Type is displayed. Press up  or down to select 'Reset'
12.
Press page  and scroll  together to return to the home display.
13.
To run the program either press RUN/HOLD or:
Press page  until Program Status is displayed
14.
Press scroll  until the cursor moves to the program status of 'Reset'
15.
Press  to select 'Run'.
16.
30min
400°C
5°C/m
Segment 1
Segment 2
Type 'Rate'
Type 'Dwell'
Target 400°
Duration
Rate 5.0°/Min
30min
MC19-1.20
30min
600°C
Segment 3
Segment 4
Type 'Step'
Type 'Dwell'
Duration
Target 600°
30min
Time display examples
2 hours
10:05.3 Min / Sec / 10
24:10:05 Hour / Min / Sec
196:10
200°C
Segment 5
Segment 6
Type 'Time'
Type 'End'
Target 200°
End type
'Reset'
Time 2 hours
Programming
th
Sec
Hour / Min
15

Advertisement