8.4.2.1. Edit filter
To get to the Edit filter menu, press
can define the results that require sediment findings. On the left
side of the screen, you find a list with all parameters found on the
test strips. You can determine a filter criterion for every param-
eter. Such a criterion is always one possible measurement value.
Thus, you can define for which results sediment findings have to
be added. (e.g. Blood – 10 / Ket – 100 / ...)
Edit filter
BLD
UBG
BIL
PRO
NIT
KET
GLU
pH
Select
You can choose the parameters either by using the scroll arrows
and the activity-button
parameter. Once you have chosen a parameter, a new list with
choices appears on the left side of the screen.
Edit filter
BLD
UBG
BIL
PRO
NIT
KET
GLU
pH
Select
The list on the right shows the different gradations for the re-
spective parameter. Choose your desired value by pressing the
respective button. The value is displayed on the right side of the
screen.
Edit filter
BLD
UBG
BIL
PRO
NIT
KET
GLU
pH
Select
If you want to delete a filter criterion you have already entered,
select the criterion in the list on the left side of the screen and
press the activity-button
the filter criteria on the right side. If you want to enter a new crite-
rion for the parameter, select the parameter again on the left side
of the screen and make a new selection.
After you have defined your filter criteria, all results that are equal
to, or exceed one of the criteria will be selected for sediment find-
ings. The activity-button
sediment
the results that match your filter criteria.
Press
to confirm and save your changes. If you leave the
menu by pressing
, none of your choices will be adopted.
40
500
®
. In the menu, you
Edit
filter
Filter :
Clear
, or by directly touching the
Select
30
100
500
Clear
Filter :
PRO
100
mg/dl
Clear
. The parameter is removed from
Clear
in the main menu gets you to all
Enter
Note: The „Enter sediment" function and the sediment filter only
list results that were measured after the function has been set /
enabled. If you activate or change the sediment function after
measurements have been performed, those measurements will
not be captured by the new filter criteria. Therefore, the previ-
ous measurements cannot be accessed via the "Enter sediment"
menu. If the filter is not activated, all results will be selected for
sediment findings. However, the memory function enables you
to enter sediment findings for all results as well (see Memory,
chapter 10).
8.4.3. Print after
The option-field
Print
which point results should be printed. By default, the choice-
button
is activated. This means all results will be
Measurement
printed immediately after the measurement has been performed,
irrespective of the need for additional sediment information.
If you want results to be printed after possible sediment findings
have been entered, press the choice-button
that is listed for additional sediment findings is now only printed
once the sediment findings have been entered. Press
confirm and save your changes. If you leave the menu by press-
ing
, none of your choices will be adopted.
8.4.4. Print sediment settings
The printer icon
in the middle of the screen allows you to
print all sediment settings. The basic settings for sediment entry,
sediment filter, as well as all defined sediment parameters will
be printed.
8.5. Language
You can choose from 8 different languages in the language menu.
Press the choice-button for the respective language.
English
Français
Polski
Press
to confirm and save your changes. If you leave the
menu by pressing
, none of your choices will be adopted.