5.2. Power supply
The instrument requires a power supply of 7,5 V DC, which is
supplied by the power pack (110/220 V, 50/60H).
Connect the power cable to the transformer, and then connect
the transformer cable to the DC IN port (a) at the back of the
instrument. Connect the cable to an outlet.
a
Note: Only operate the instrument with the included, original
power pack!
Note: Only start the instrument once the transport rack has been
inserted and all power connections are complete.
5.3. Connections / interface
Apart from the power connection, the instrument has 4 additional
interfaces (1 x USB (b), 1 x PS/2 (c), 2 x RS232 (d)) to connect
different devices or a laboratory information system (LIS) (see
above).
5.4. Insert printer paper
To open the printer paper cover, press the little button above the
touch screen. Move the printer paper cover backwards.
Unroll appr. 10cm (4") of printer paper and place the printer pa-
per into the chamber, so that the paper unrolls downwards and
towards the touch screen.
Manual URYXXON
MN
b
c
d
®
500 V 01 / 03.10
Manual URYXXON
Close the printer paper cover with an audible click, so that the
printer paper comes out from the chamber.
Note: Only use special thermal printer paper!
Switching on the instrument
You can find the On / Off switch (e) on the back of the instru-
ment. Turn on the instrument. Right after switching on, the instru-
ment performs a self-test. The touch screen lights up and the
start screen appears. The little light above the transport tray turns
green, thus signalling operational readiness.
e
500
®
35