Sealey TA101 Instructions page 2

12 function digital automotive analyser

Advertisement

3. SPECIFICATION
AC voltage (Autorange)
Range
Resolution
3.26V
1mV
32.6V
10mV
326V
0.1V
750V
1V
Input Impedance: 10MΩ
Frequency range: 40Hz to 400Hz
Overload Protection: DC 1000V; AC 750V.
DC voltage (Autorange)
Range
Resolution
326mV
0.1mV
3.26V
1mV
32.6V
10mV
326V
0.1V
1000V
1V
Input Impedance: 10MΩ (for 326mV range: >100MΩ)
Overload Protection: DCV1000V; ACV750V.
AC Current (µA and mA are Autorange)
Range
Resolution
326µA
0.1µA
3260µA
1µA
32.6mA
10µA
326mA
0.1mA
10A
10mA
Overload Protection: F 500mA L 250V
(for 10A range: F10A L 250V, Maximum measurement duration
<10 secs. and interval >15 mins). Frequency: 40Hz to 400Hz.
DC Current (µA and mA are Autorange)
Range
Resolution
326µA
0.1µA
3260µA
1µA
32.6mA
10µA
326mA
0.1mA
10A
10mA
Overload Protection: F 500mA L 250V
(for 10A range: F10A L 250V, Maximum measurement duration
<10 secs. and interval >15 mins).
Duty Cycle test
Range
Scope (%)
Resolution
Duty
0~99.9
4. OPERATION
WARNING! Ensure that you read, understand and apply the safety and operational instructions before connecting the analyser. Only when you are sure
that you understand the procedures is it safe to proceed with testing.
WARNING! risk of electrocution. High voltage circuits, both AC and DC are very dangerous and should be measured with great care.
Operating temperature range 0°C to 40°C.
Remember to turn off analyser when measurement is completed.
Note:
If "OL" appears in the display during a measurement, the value exceeds the range you have selected. Select a higher range.
Note:
On some low AC and DC ranges, with the test leads not connected to a device, the display may show a random fluctuating reading. this is
normal and is caused by the high input sensitivity. the reading will stabilise and give a correct measurement when connected to a circuit.
4.1.
Data Hold Button
4.1.1.
the data hold function allows the analyser to freeze a measurement reading for later reference.
4.1.2.
Press the data hold button once to freeze the reading in the display. the indicator "H" will appear in the display.
4.1.3.
Press the data hold button again to return to normal operation.
4.2.
AC/DC voltage measurement
4.2.1.
Insert the black test lead into the negative COm socket and the red test lead into the positive
4.2.2.
turn the function dial to the voltage setting and press the function selection button to select AC or DC as required.
4.2.3.
Connect the test leads to the circuit under test and read the voltage on the display.
4.2.4.
For DC voltage, the polarity of the red test lead will be displayed as well as the value.
4.3.
DC Current measurement
WARNING! Do not make current measurements at 10A for longer than 10 seconds in every 15 minutes. Exceeding this may cause damage to
the analyser and test leads.
4.3.1.
Insert the black test lead into the negative COm socket, and the red test lead into the positive µAmA socket (or the positive 10A socket for currents from
326mA to 10A).
4.3.2.
turn the function dial to the appropriate setting as required (see ranges above). If the current range to be measured is not known, set to the highest
range setting, and then select to the correct range when the first reading is taken, until satisfactory resolution is obtained. Press the function selection
button to select AC or DC as required.
4.3.3.
Connect the test leads in series with the circuit under test and read the current on the display.
4.3.4.
For DC current, the polarity of the red test lead will be displayed as well as the value.
Accuracy
± (0.8% + 5) (40~200Hz)
± (1.2% + 7) (200~400Hz)
± (1.0% + 5) (40~200Hz)
± (1.5% + 7) (200~400Hz)
Accuracy
± (0.5% + 5)
± (0.8% + 5)
± (1.0% + 8)
Accuracy
± (1.5% + 5) (40~200Hz)
± (1.8% + 7) (200~400Hz)
± (3.0% + 5)
Accuracy
± (1.2% + 5)
± (2.0% + 5)
Accuracy
0.1
± (2.0% + 5)
Dwell Angle (manual Range)
Cylinder Range
2CYL
3CYL
4CYL
5CYL
6CYL
8CYL
Frequency
Range
Accuracy
320Hz
3200Hz
32kHz
Tachometer
Range
10 x rPM
Temperature
-20°C~750°C
-4°F~1400°F
Audible Continuity Test
If the resistance of the circuit
under test is lower than 50Ω,
an audible warning will sound.
Diode test
the approximate forward
voltage of the diode under test
will be displayed on the LCD.
Resistance (Autorange)
Range
326Ω
3.26kΩ
32.6kΩ
326kΩ
3.26MΩ
32.6MΩ
Original Language Version
Resolution
0 ~180.0°
0 ~120.0°
0 ~ 90.0°
0.1°
0 ~ 72.0°
0 ~ 60.0°
0 ~ 45.0°
(manual
Range)
Resolution
0.1Hz
1Hz
0.1kHz
Scope (RPm)
Resolution
rPM
0~3260
1 rPM
10 x (0~3260)
10 rPM
Range
Resolution
-20~0°C(-4°F~32°F) ±(6.0% +5)
1°C
0~400°C(32°F~752°F) ±(1.5% +5)
1°F
400~750°C(752°F~1400°F) ±(1.8%+5)
Description
Description
Resolution
0.1Ω
1Ω
10Ω
0.1kΩ
1kΩ
10kΩ
v
RPm
socket.
Accuracy
± (2.5% + 5)
± (2.5% + 5)
Accuracy
± (2.5% + 5)
Accuracy
test condition
Open circuit voltage is
approximately 3V.
test condition
the forward DC current is
approximately 1mA, the
reversed DC voltage is
approximately 3V.
Accuracy
± (1.0% + 8)
± (1.0% + 5)
± (3.0% + 7)
tA101 Issue No.2 - 19/12/11