DPD-4
(Total Chlorine or Ozone)
CL1
REMOVE STRIP
1
®
Remove one (1) eXact
Strip Micro CL (DPD-4 for Total Chlorine or Ozone), part number
486670, from the bottle before beginning the test. Set the strip in a dry, convenient place and
recap the bottle immediately.
TURN METER ON
2
Press the ZERO/ON button to power the meter on; the display will show all annunciators,
then the current MENU selection, followed by the last reading.
SELECT TEST: CL1
3
Press and re-press the MENU button until the display shows the parameter you want to run.
FILL METER WITH SAMPLE
4
Rinse the CELL at least 3 times with the water sample you will be testing - rinsing
minimizes the potential for cross- contamination from a previous test. Finally, fill cell to
capacity (4ml) with the water sample.
ZERO METER
5
Press the ZERO/ON button. The cursor will move across the display, followed by 0.00 PPM.
Sample is ready for testing.
DIP STRIP AND PRESS "READ"
6
Dip the eXact
®
Strip Micro CL (DPD-4), Part No. 486670 into the CELL and immediately press
READ. This starts the 20 SECOND countdown timer. During this time move the strip in a gentle
back and forth motion. Remove and discard the strip when the time displays 1. The cursor will
move across the display, informing you that it is about to measure the sample. Record result
displayed (this result is automatically stored in CL1).
PRESS READ AGAIN
7
Press READ again and the meter will count down and display the next Total Chlorine result.
If this reading matches the previous result, then record this as Total Chlorine result. This
result is stored in CL1. If reading has increased by 0.01, press READ again and repeat.
After testing is completed, rinse cell immediately. Record the Total Chlorine as the highest value
the meter displayed.
NOTE: Standard Method (4500-Cl G, procedure for total chlorine) requires the reading to be
made 2 minutes after the reagents are added.
pH Test Procedure
PH2
REMOVE STRIP
1
Remove one (1) eXact
®
Strip Micro PH, part number 486639, from the bottle before
beginning the test. Set the strip in a dry, convenient place and recap the bottle immediately.
TURN METER ON
2
Press the ZERO/ON button to power the meter on; the display will show all annunciators,
then the current MENU selection, followed by the last reading.
SELECT TEST: PH2
3
Press and re-press the MENU button until the display shows the parameter PH2.
FILL METER WITH SAMPLE
4
Rinse the CELL at least 3 times with the water sample you will be testing - rinsing minimizes
the potential for cross-contamination from a previous test. Finally, fill cell to capacity (4ml) with
the water sample.
ZERO METER
5
Press the ZERO/ON button. When the display shows 0.0 PH, the sample is ready for testing.
DIP STRIP AND PRESS "READ"
6
®
Dip the eXact
Strip Micro PH, Part No. 486639 into the CELL and immediately press READ.
This starts the 20 SECOND countdown timer. During this time move the strip in a gentle back
and forth motion. Remove and discard the strip when the time displays 1. The cursor will move
across the display, informing you that it is about to measure the sample. Record result displayed
(this result is automatically stored in PH2). After testing is completed, rinse cell immediately.
NOTE: For best results, Total Alkalinity of the sample should be 40-140 ppm.
PH2: pH reacts with Phenol Red, which is yellow at pH 6.1, but forms a red color as the pH level
rises. Maximum red is at about pH 8.4.
CL
Test Procedure
PH
6
NOTES
27