Download Print this page

Industrial Test Systems 486691 Instruction Manual page 2

Ideal for drinking water, pools and spas, environmental, & educational testing

Advertisement

eXact
eX
Micro 7+ Photometer
®
B
A
(Used for Cyanuric
Acid, Fluoride, Iron,
and Manganese tests)
eXact
Micro 7+ Meter Specifications
cro
®
Measurement Method:
Photometric
Light Source:
Light Emitting Diode (LED)
Wavelength:
525 nm
Transmission Range:
100 - 0.00 %T
Photometric Precision:
+/- 0.1/0.01 %T
Automatic Range Selection:
See Specifications below
Display:
3-digit customized liquid crystal display
with annunciators
CELL Pathlength:
20mm
We offer a "Green" Alternative
®
eXact
Strip Micro 7+ has been designed to offer the user a more "Green" and cost-effective alternative to testing.
®
Instead of using a 10ml water sample, eXact
Strip Micro 7+ uses a 4ml water sample, which uses up to 60% less
chemical per test. The accuracy of the meter is maintained by designing the photo cell with a 20mm pathlength.
eXact
®
Micro 7+ Specifications
Menu Tests for
Range
CL1
Free Chlorine (DPD-1)
0.00-2.39 ppm
& Total Chlorine (DPD-3) 2.4-6.0 ppm
PH2
pH
6.2-8.4 pH
BR3
Bromine (DPD-1)
0.00-2.99 ppm
3.0 - 9.0 ppm
AL4
Total Alkalinity
20-180 ppm
CA5
Calcium as CaCO
20-990 ppm
3
CU6
Copper (Cu
+2
)
0.00-1.99 ppm
2.0 - 8.0 ppm
TR7
Transmission
99.9-10.0 %T
9.99-0.01 %T
(17 other test parameters)
ITEM PICTURED
COMPONENT
LETTER
NAME
A
Mixing Cap
B
Cell (Built-in Plastic, 4 ml)
C
C
LCD Display
D
READ Button
E
MENU Button
F
ZERO/ON Button
D
G
Battery Cover with loop for lanyard
H
AAA Batteries (x4)
E
F
H
ter Specifications
Cell Chamber:
Custom-molded, proprietary, PET plastic
fused into chamber, non-removable
Sample Required:
4 ml (0.13 oz)
Operating Temperature Range:
0 - 50°C (32° - 122°F)
Power Supply:
(4) AAA alkaline batteries
Battery Life:
>2000 tests with alkaline batteries
Electromagnetic Compliance:
Emitted Interference - EN 61326
(EMC)
Immunity to Interference - EN 61326
Waterproof Rating:
Exceeds IP67
Weight:
Instrument: 140 g (5 oz)
Dimensions:
Instrument: 5 (W) x 3.5 (D) x 16.5 (H) cm;
(2 x 1.4 x 6.375 in)
Resolution
+/- Accuracy
0.01
0.02
0.1
0.1, or 4%
0.1
0.3
0.01
0.03
0.1
0.1, or 4%
1
25
10
20 or 7%
0.01
0.02
0.1
0.1 or 4%
0.1
0.1
0.01
0.01
2
eXact
Strip Micro 7+ Reagent Reorder Information
®
G
+
* Results utilize the Tr-7 (Transmission) meter function and require the use of a conversion table. See respective test procedures
for more information and tables.
** Total Chlorine DPD-3 Test requires Free Chlorine DPD-1 (486637) to be run first.
PRICES & SPECIFICATIONS SUBJECT TO CHANGE WITHOUT NOTICE.
NOTE: Because most of our products are test strips or use reagents
that have little or no hazard in the quantity sold, MSDS sheets are
not supplied with the test. The exceptions are the Manganese
(486606) test, which comes with 2 strips and one liquid reagent
(PAN); Fluoride (486643) test, which is a liquid reagent (SPADNS),
and Iron (481623) test, which is a powder reagent.
If your required procedure is not listed in this manual, please see the back page
for our contact information.
To ensure optimal performance, store your eXact
excess heat (below 100°F / 38°C), moisture, and oxidizers such as Chlorine and Bromine.
®
kit in a cool, dry place away from
31

Advertisement

loading

This manual is also suitable for:

Exact micro 7+