Download Print this page

Industrial Test Systems 486691 Instruction Manual page 13

Ideal for drinking water, pools and spas, environmental, & educational testing

Advertisement

For Pool and Spa Salt Systems
Chloride (as NaCl) Test Procedure
TR7
TURN METER ON
1
Press the ZERO/ON button to power the meter on; the display will show all annunciators,
then the current MENU selection, followed by the last reading.
SELECT TEST: TR7
2
Press and re-press the MENU button until the display shows the parameter TR7.
PREPARE THE CELL
3
Using deionized or distilled water, rinse the cell at least 3 times to minimize potential cross
contamination from a previous test. Finally, fill the cell to the top with deionized or distilled water.
ADD THE SAMPLE
4
Use the small pipet provided to add 0.1ml of the sample to the water in the cell.
ZERO METER
5
Press the ZERO/ON button. The cursor will move across the display, followed by 100 %T.
Sample is ready for testing.
DIP STRIP
6
Dip the eXact
®
Strip Micro CHLORIDE (as NaCl), Part No. 481657 into the CELL and
immediately press READ. This starts the 20 SECOND countdown timer. During this time move
the strip in a gentle back and forth motion. Remove and discard the strip when the time
displays 1. The cursor will move across the display, informing you that it is about to measure
the sample. Record result displayed (this result is automatically stored in TR7).
USE TABLE
7
Find the "TR" result in the table below to determine the Sodium Chloride concentration in ppm
(parts per million). (Example: a "TR" result of 65.3 (use only the 65 for the chart) equals a Sodium
Chloride value of 4600 ppm). Record result. After testing is completed, rinse cell immediately.
Chloride
Sodium Chloride results require the table below. Follow eXact
®
using eXact
Strip Micro CHLORIDE, Part No. 481657. NOTE: To convert the NaCl value to Chloride (Cl
multiply the value from the chart below by 0.6. (Example: 4600ppm NaCl = 2760ppm Cl
®
eXact
Strip Micro CHLORIDE, Part No. 481657 - for 4mL Samples
tr
9
8
7
6
<100
<100
100
200
90
80
1300
1500
1700
1800
70
2900
3000
3100
3300
4200
4300
4400
4500
60
50
5300
5400
5500
5600
6500
6600
40
6300
6400
30
7200
7300
7400
7500
20
8200
8300
8400
8500
9200
9400
9500
9600
10
0
>11000
>11000
>11000
>11000
Table
(as NaCl)
®
Micro 7+ Chloride (as NaCl) Test Procedure (above)
5
4
3
2
1
400
600
700
900
1000
2000
2100
2300
2400
2600
3500
3700
3800
3900
3400
4600
4700
4800
5000
5100
5700
5800
5900
6000
6100
6700
6700
6800
6900
7000
7600
7700
7800
7900
8000
8600
8700
8800
8900
9000
9700
9800
10000
10200
10500
>11000
>11000
>11000
>11000
>11000
Rev. 121608 NACL
20
Chloride (as NaCl) Test Procedure
CHLORIDE
CHLORIDE
TURN METER ON
1
Press the ZERO/ON button to power the meter on; the display will show all annunciators,
then the current MENU selection, followed by the last reading.
SELECT TEST: TR7
2
Press and re-press the MENU button until the display shows the parameter TR7.
FILL METER WITH SAMPLE
3
Rinse the CELL at least 3 times with the water sample you will be testing - rinsing minimizes
the potential for cross-contamination from a previous test. Finally, fill cell to capacity (4ml)
with the water sample.
ZERO METER
4
Press the ZERO/ON button. The cursor will move across the display, followed by 100 %T.
Sample is ready for testing.
DIP STRIP AND PRESS "READ"
5
Dip the eXact
immediately press READ. This starts the 20 SECOND countdown timer. During this time move
the strip in a gentle back and forth motion. Remove and discard the strip when the time
displays 1. The cursor will move across the display, informing you that it is about to measure
the sample. Record result displayed (this result is automatically stored in TR7).
USE TABLE
6
Find the "TR" result in the table below to determine the Chloride concentration in ppm (parts per
million). (Example: a "TR" result of 65.3 (use only the 65 for the chart) equals a Chloride value of
115 ppm). Record result. After testing is completed, rinse cell immediately.
Sodium Chloride results require the table below. Follow eXact
Procedure (above) using eXact
Chloride (Cl
), multiply the value from the chart below by 0.6. (Example: 115ppm NaCl = 69ppm Cl
),
)
eXact
tr
9
90
<3
0
33
80
1200
70
73
2700
60
105
108
4000
133
135
50
5200
40
158
160
6200
30
180
183
7100
205
208
20
8100
10
230
235
9100
0
285
>290
11000
>11000
To determine NaCl concentration from 100-11000 ppm for salt pools, do a 1 to 40 dilution of the pool water
and perform test as described on page 20.
For Tap Water
®
Strip Micro CHLORIDE (as NaCl), Part No. 481657 into the CELL and
Chloride (as NaCl) Table
®
Strip Micro 7+ Chloride (as NaCl) Test
®
Strip Micro CHLORIDE, Part No. 481657. NOTE: To convert the NaCl value to
®
Strip Micro CHLORIDE, Part No. 481657 - for 4mL Samples
8
7
6
5
4
<3
3
5
10
15
38
43
45
50
53
75
78
83
85
88
110
113
115
118
138
140
143
145
163
165
168
168
185
188
190
193
210
213
215
218
238
240
243
245
>290
>290
>290
>290
13
TR7
)
3
2
1
0
18
23
25
30
58
60
65
68
93
95
98
100
120
125
128
130
148
150
153
155
170
173
175
178
203
195
198
200
220
223
225
228
250
255
263
273
>290
>290
>290
>290
Rev. 121608 NACL