Brew Calibration Instructions - Marco Qwikbrew 2 Service Manual

Version 2, 2004 onwards boiler brewer
Hide thumbs Also See for Qwikbrew 2:
Table of Contents

Advertisement

4.3 Brew calibration Instructions:

Setting Brew Water Volumes:
When in volume calibration mode you are selecting the overall water flow time for each
brew. The spray head outputs approx. 700mL/minute which is 11.66ml/second. You can
select 1, 2 or all 3 brew volumes in the one calibration session, they can be selected in any
order, e.g. brew 2 could be the lowest or highest volume depending on when you press the
brew 2 button. So if you want a volume of 3.5L in brew 2 you would press 2 after 5min
(=300sec), (11.66mlx300s=3.5L), see table guide.
To Calibrate Volume:
1. Turn the brewer on by pressing the power button.
2. Press and hold any of the brew buttons for 5+ seconds, your timer should be started
the moment you press the button
calibration mode, e.g. you could press & hold brew 1 if only calibrating brew 3)
3. The power on red LED will blink to show you are in volume calibration mode, no water
will flow at this time.
4. When the desired time has passed as per table 1 below press the desired brew button
1,2,3 to select the corresponding volume for that brew.
5. When all required brews are selected turn the brewer off using the power button to
save your settings.
Service manual April 2017 Qwikbrew 2 version 2
down. (any brew button can be pressed to enter
Page 9 of 26